Model 400E Ozone Analyzer
Instruction Manual
Getting Started
P/N 04316 Rev: B
29
3.1.2.
Pneumatic Connections:
CAUTION
OZONE (O3) IS A TOXIC GAS.
Obtain a Material Safety Data Sheet (MSDS) for this material.
Read and rigorously follow the safety guidelines described
there.
Do not vent calibration gas and sample gas into enclosed areas.
1.
Sample and calibration gases should only come into contact with PTFE (Teflon),
FEP, glass, stainless steel or brass.
Dry Air
Inlet
Zero Air
Inlet
Span Gas
Inlet
Status
Outputs
Serial I/O
LED’s
Control
Inputs
Analog
Outputs
COM B Connector
(RS
-
232 or RS
-
485)
Exhaust
Outlet
Sample
Inlet
Cooling
Fan
AC Power
Receptacle
COM A Connector
(RS232 Only)
DCE
–
DTE
Switch
Figure 3-2: Rear Panel Connectors
CAUTION
In order to prevent dust from getting into the gas flow channels
of your analyzer, it was shipped with small plugs inserted into
each of the pneumatic fittings on the back panel.
Make sure that all of these dust plugs are removed before
attaching exhaust and supply gas lines.