Configuring the Internal Zero/Span Option
(IZS)
In order to use the IZS option to perform calibration checks, it is necessary to
configure certain performance parameters of the O
3
Generator.
6.4.1.
Setting the O
3
Generator Span-Check Output Level
To set the ozone SPAN concentration for the IZS O
3
generator, press:
Toggle these
keys to
change
setting
EXIT
ignores the
new setting.
ENTR
accepts the
new setting.
SAMPLE RANGE = 500.0 PPB O3 =XXX.X
< TST CAL CALZ
CALS
SETUP
SPAN CAL M
RANGE = 500.0 PPM
O3 =XXX.XX
<TST TST> SPAN
CONC
EXIT
SPAN CAL M
SPAN CONC MENU
O3 =XXX.XX
SPAN
O3GN
EXIT
SPAN CAL M
3) O3_GEN_SET=
400.0 PPB
0 4 0 0 .0 ENTR
EXIT
•
Only Values from 0 to 1500 will be accepted..
•
A value of 0 turns the lamp OFF.
•
The
ENTR
key will disappear if an invalid setting is
attempted.