Operating Instructions
Model 400E Ozone Analyzer
Instruction Manual
58
P/N 04316 Rev: B
To View and Clear the various warning messages press:
SAMPLE
WHEEL TEMP WARNING
O3 = XXX.X
TEST
CAL
MSG
CLR
SETUP
Make
SURE
warning
messages are
NOT
due to
LEGITIMATE PROBLEMS.
.
Press
CLR
to clear the
message currently being
Displayed.
If more than one warning is
active the next message will
take its place
Once the last warning has been
cleared, the analyzer returns to
SAMPLE
Mode
SAMPLE
RANGE=500.0 PPB
O3 = XXX.X
< TST TST > CAL
MSG
CLR
SETUP
SAMPLE
WHEEL TEMP WARNING
O3 = XXX.X
TEST CAL
MSG
CLR
SETUP
TEST
deactivates Warning
Messages
MSG
activates Warning
Messages.
<TST TST
> keys replaced with
TEST
key
6.2.2.
Test Functions
A series of Test Functions are available for viewing via the Front Panel display while
the analyzer is in
SAMPLE
mode. These functions provide valuable information
regarding the present operation and status of the instrument. These parameters
are also useful during troubleshooting exercises (see Section 11.1.2).
Table 6-2: Test Functions Defined
Parameter
Display Title
Units
Meaning
RANGE
RANGE - -
RANGE1
RANGE2
PPB, PPM,
UGM & MGM
The Full Scale limit at which the output range of
the analyzer’s ANALOG OUTPUTS are currently
set.
THIS IS NOT
the Physical Range of the
instrument. See Section 6.4 for more information.
If
DUAL
or
AUTO
Range modes have been
selected, two
RANGE
functions will appear, one
for each range.
STABILITY
STABIL
mV
Standard deviation of O
3
Concentration readings.
Data points are recorded every ten seconds. The
calculation uses the last 25 data points.
PHOTOMEAS
O3 MEAS
mV
The average UV Detector output during the
SAMPLE portion of the analyzer’s measurement
cycle.
PHOTOREF
O3 REF
mV
The average UV Detector output during the
REFERENCE portion of the analyzer’s
measurement cycle.
O3GENREF
O3 GEN
(2)
mV
The current output of the O
3
generator reference
detector representing the relative intensity of the
O
3
generator UV Lamp.
(2)