TROUBLESHOOTING & REPAIR PROCEDURES
Model 400E Ozone Analyzer
Instruction Manual
268
P/N 04316 Rev: B
11.6.3.2.
Metal Wool Scrubber Option
Contact T-API for instructions on replacing the optional Metal Wool Scrubber.
11.6.4.
Replacing the IZS O
3
Scrubber
1.
Turn off power to the instrument.
2.
Remove instrument cover.
3.
The IZS Zero Air Scrubber is attached to the brass elbow inlet fitting on the top
of the O
3
Generator Assembly. See Figure 11-6.
4.
Disconnect 1/4” Tube Fitting nut on O3 Generator inlet fitting.
5.
Disconnect 1/8” tube fitting on the other end of the scrubber.
6.
Install new scrubber by reversing these steps.
IZS Zero Air Scrubber
Figure 11-6: IZS Zero Air Sc rubber Location