Model 200E Instruction Manual
Operating Instructions
6.7. Setup - Diagnostics (DIAG)
A series of diagnostic tools is grouped together under the
SETUP-MORE-DIAG
menu. As
these parameters are dependent on firmware revision, the menu trees are listed in detail in
Appendix A. The individual parameters, however, are explained in more detail in the
following section, indicated in Table 6-6. These tools can be used in a variety of trouble-
shooting and diagnostic procedures and are referred to in many places of the maintenance
and trouble-shooting sections.
Table 6-6: M200E Diagnostic (DIAG) Functions
Diagnostic Function and Meaning
Front Panel
Mode Indicator
Section
SIGNAL I/O:
Allows observation of all digital and analog signals in
the instrument. Allows certain digital signals such as valves and
heaters to be toggled ON and OFF.
DIAG I/O
ANALOG I/O:
When entered, the analyzer performs an analog output
step test. This can be used to calibrate a chart recorder or to test the
analog output accuracy.
DIAG AOUT
ANALOG I/O
CONFIGURATION:
Analog input/output parameters
are available for viewing and configuration.
DIAG AIO
TEST CHAN OUTPUT:
Configures the
A4
analog output channel.
DIAG TCHN
OPTIC TEST:
When activated, the analyzer performs an optic test,
which turns on an LED located inside the sensor module near the PMT
(Fig. 10-15). This diagnostic tests the response of the PMT without
having to supply span gas.
DIAG OPTIC
ELECTRICAL TEST:
When activated, the analyzer performs an
electric test, which generates a current intended to simulate the PMT
output to verify the signal handling and conditioning of the PMT
preamp board.
DIAG ELEC
OZONE GEN OVERRIDE:
Allows the user to manually turn the O
3
generator on or off. This setting is retained when exiting
DIAG
.
DIAG OZONE
FLOW CALIBRATION:
This function is used to calibrate the gas flow
output signals of sample gas and ozone supply. These settings are
retained when exiting
DIAG
.
DIAG FCAL
044100102 Rev A
57
Summary of Contents for 200E
Page 15: ...M200E Documentation Model 200E Instruction Manual USER NOTES xiv 044100102 Rev A ...
Page 19: ...M200E Documentation Model 200E Instruction Manual USER NOTES 4 044100102 Rev A ...
Page 141: ...Calibration Procedures Model 200E Instruction Manual USER NOTES 126 044100102 Rev A ...
Page 203: ...Theory of Operation Model 200E Instruction Manual USER NOTES 188 044100102 Rev A ...
Page 239: ...Troubleshooting Repair Model 200E Instruction Manual USER NOTES 224 044100102 Rev A ...
Page 265: ...APPENDIX A Model 200E Instruction Manual USER NOTES 250 044100110 Rev A ...
Page 271: ...APPENDIX B Model 200E Instruction Manual USER NOTES 256 044100120 Rev A ...
Page 275: ...APPENDIX D Model 200E Instruction Manual 260 044100141 Rev A USER NOTES ...