Calibration Procedures
Model 200E Instruction Manual
•
Perform the zero/span check. Zero and span checks using the zero/span valve or
IZS option are similar to that described in Section 7.3, except:
On units with an IZS option installed, zero air is supplied to the analyzer through the zero
gas inlet and the same gas stream is used to create span gas internally using the permea-
tion device.
On units with a zero/span valve option installed, zero air and span gas are supplied to the
analyzer through the zero gas and span gas inlets from two different sources.
The zero and calibration operations are initiated directly and independently with dedicated
keys
CALZ and CALS
.
Note that there is no option to use a dedicated span gas port if the IZS option is installed.
Span gas from a gas tank would need to be supplied through the Sample port or through
one of the auxiliary ports available on the rear panel (with modified internal plumbing),
although the user could modify the plumbing to have the span gas go through the IZS oven
(without the permeation tube installed).
W ait until
NOX
STB
falls below
0.5 ppb. This may
take several
minutes.
The value of
NOX
STB
may jump
significantly. Wait
until
NOX STB
falls
below
0.5 ppb. This
may take several
minutes.
SPAN CAL M
NOX STB=XXX.X PPB
NOX=X.XX
< TST TST >
ZERO
SPAN
CONC
EXIT
SAMPLE
NOX STB=XXX.X PPB
NOX=X.X
< TST TST > CAL
CALZ
CALS
SETUP
ZERO CAL M
NOX STB=XXX.X PPB
NOX=X.X
< TST TST >
ZERO
CONC
EXIT
SAMPLE
NOX STB=XXX.X PPB
NOX=X.X
< TST TST > CAL
CALZ
CALS
SETUP
EXIT
returns to the main
SAMPLE
display
ACTION:
Record the
NO
X
, NO
2
& NO
readings presented in
the upper right corner
of the display.
ACTION:
Record the
NO
X
, NO
2
& NO
readings presented in
the upper right corner
of the display.
SAMPLE RANGE = 500.0 PPB
NOX=X.XX
< TST TST >
CAL
CALZ
CALS
SETUP
Scroll to the
NOX
STB
test function.
7.7. Calibration With Independent or AUTO Ranges
Additional considerations apply when calibrating a Model 200E with reporting ranges set to
AUTO
(automatic) or
IND
(independent) mode (Sections 6.5.4 and 6.5.5).
7.7.1. Calibration in AUTO Range Mode
If the analyzer is operated in
AUTO
range mode, the high and low ranges must be indepen-
dently calibrated. After pressing the
CAL
,
CALZ
or
CALS
keys, the user is prompted for the
range to be calibrated as seen in the zero calibration example below.
120
044100102 Rev A
Summary of Contents for 200E
Page 15: ...M200E Documentation Model 200E Instruction Manual USER NOTES xiv 044100102 Rev A ...
Page 19: ...M200E Documentation Model 200E Instruction Manual USER NOTES 4 044100102 Rev A ...
Page 141: ...Calibration Procedures Model 200E Instruction Manual USER NOTES 126 044100102 Rev A ...
Page 203: ...Theory of Operation Model 200E Instruction Manual USER NOTES 188 044100102 Rev A ...
Page 239: ...Troubleshooting Repair Model 200E Instruction Manual USER NOTES 224 044100102 Rev A ...
Page 265: ...APPENDIX A Model 200E Instruction Manual USER NOTES 250 044100110 Rev A ...
Page 271: ...APPENDIX B Model 200E Instruction Manual USER NOTES 256 044100120 Rev A ...
Page 275: ...APPENDIX D Model 200E Instruction Manual 260 044100141 Rev A USER NOTES ...