Troubleshooting & Repair
Model 200E Instruction Manual
•
Attach the outlet port of a flow meter to the sample inlet port on the rear panel.
Ensure that the inlet to the flow meter is at atmospheric pressure.
•
The sample flow measured with the external flow meter should be 500 cm³/min
±
10%. If a combined sample/ozone air Perma Pure dryer is installed (optional equip-
ment), the flow will be 640 cm³/min ± 10% (500 cm³/min for the sample and 80
cm³/min for the ozone generator supply air and 60 cm³/min for the purge flow).
•
Low flows indicate blockage somewhere in the pneumatic pathway.
11.5.4. AC Power Configuration
The M200E can be easily configured for two main power regimes, 100-120 V and 220-240 V
at either 50 or 60 Hz. The analyzer is correctly configured for the AC power voltage in use if
the voltage selector plug is configured properly. See Figure 3-1 and Figure 11-4 for the
location of this plug. The plug for 100-120 V should have white jumper wires, the plug for
220-240 V should have blue jumper wires. Note, that an analyzer configured for 230 V will
still turn on at 115 V (and vice versa), but the heaters may burn out or not heat up fast
enough. Internally, several LEDs should turn on as soon as the power is supplied. If an
incorrect power configuration is suspected, check for the correct voltage and frequency at
the line input on the rear panel.
Note that the analyzer will be severely damaged if 220-240 V is supplied to it when
configured for 100-120 V. Never bypass the power switch or circuit breaker.
11.5.5. DC Power Supply
If you have determined that the analyzer’s AC main power is working, but the unit is still
not operating properly, there may be a problem with one of the instrument’s switching
power supplies, which convert AC power to 5 and ±15 V (PS1) as well as +12 V DC power
(PS2). The supplies can either have no DC output at all or a noisy output (fluctuating).
To assist tracing DC Power Supply problems, the wiring used to connect the various printed
circuit assemblies and DC powered components and the associated test points on the relay
board follow a standard color-coding scheme as defined in Table 11-3.
Table 11-3: DC Power Test Point and Wiring Color Code
Name Test
Point#
Color
Definition
DGND 1
Black
Digital
ground
+5V 2
Red
AGND 3
Green
Analog
ground
+15V 4 Blue
-15V 5
Yellow
+12R
6
Purple
12 V return (ground) line
+12V 7
Orange
A voltmeter should be used to verify that the DC voltages are correct as listed in
Table 11-5. An oscilloscope, in AC mode and with band limiting turned on, can be used to
evaluate if the supplies are excessively noisy (>100 mV peak-to-peak).
206
044100102 Rev A
Summary of Contents for 200E
Page 15: ...M200E Documentation Model 200E Instruction Manual USER NOTES xiv 044100102 Rev A ...
Page 19: ...M200E Documentation Model 200E Instruction Manual USER NOTES 4 044100102 Rev A ...
Page 141: ...Calibration Procedures Model 200E Instruction Manual USER NOTES 126 044100102 Rev A ...
Page 203: ...Theory of Operation Model 200E Instruction Manual USER NOTES 188 044100102 Rev A ...
Page 239: ...Troubleshooting Repair Model 200E Instruction Manual USER NOTES 224 044100102 Rev A ...
Page 265: ...APPENDIX A Model 200E Instruction Manual USER NOTES 250 044100110 Rev A ...
Page 271: ...APPENDIX B Model 200E Instruction Manual USER NOTES 256 044100120 Rev A ...
Page 275: ...APPENDIX D Model 200E Instruction Manual 260 044100141 Rev A USER NOTES ...