APPENDIX A
APPENDIX A-3:
Warnings and Test Measurements, Revision C.8
Table A-2: M200E Warning Messages, Revision C.8
Name Message
Text
Description
WSYSRES
SYSTEM RESET
Instrument was power-cycled or the CPU was reset.
DATA INITIALIZED
Data storage was erased.
WCONFIGINIT
CONFIG INITIALIZED
Configuration storage was reset to factory configuration or
erased.
WSAMPFLOW
SAMPLE FLOW WARN
Sample flow outside of warning limits specified by
SFLOW_SET
variable.
WOZONEFLOW OZONE
FLOW
WARNING
Ozone flow outside of warning limits specified by
OFLOW_SET
variable.
WOZONEGEN
OZONE GEN OFF
Ozone generator is off. This is the only warning message that
automatically clears itself. It clears itself when the ozone
generator is turned on.
WRCELLPRESS
RCELL PRESS WARN
Reaction cell pressure outside of warning limits specified by
RCELL_PRESS_SET
variable.
WBOXTEMP
BOX TEMP WARNING
Chassis temperature outside of warning limits specified by
BOX_SET
variable.
RCELL TEMP WARNING Reaction cell temperature outside of warning limits specified by
RCELL_SET
variable.
4
MANIFOLD
TEMP
WARN
Bypass or dilution manifold temperature outside of warning limits
specified by
MANIFOLD_SET
variable.
WIZSTEMP
IZS TEMP WARNING
IZS temperature outside of warning limits specified by
IZS_SET
variable.
WCONVTEMP
CONV TEMP WARNING
Converter temperature outside of warning limits specified by
CONV_SET
variable.
WPMTTEMP
PMT TEMP WARNING
PMT temperature outside of warning limits specified by
PMT_SET
variable.
WAUTOZERO
AZERO WRN XXX.X MV Auto-zero reading above limit specified by
AZERO_LIMIT
variable. Value shown in message indicates auto-zero reading at
time warning was displayed.
WHVPS
HVPS WARNING
High voltage power supply output outside of warning limits
specified by
HVPS_SET
variable.
WDYNZERO
CANNOT DYN ZERO
Contact closure zero calibration failed while
DYN_ZERO
was set
to
ON
.
WDYNSPAN
CANNOT DYN SPAN
Contact closure span calibration failed while
DYN_SPAN
was set
to
ON
.
REAR BOARD NOT DET Rear board was not detected during power up.
WRELAYBOARD
RELAY BOARD WARN
Firmware is unable to communicate with the relay board.
WFRONTPANEL
FRONT PANEL WARN
Firmware is unable to communicate with the front panel.
WANALOGCAL ANALOG
CAL
WARNING
The A/D or at least one D/A channel has not been calibrated.
WDATAINIT
WRCELLTEMP
WMANIFOLDTEMP
WREARBOARD
1
Optional equipment only
044100110 Rev A
241
Summary of Contents for 200E
Page 15: ...M200E Documentation Model 200E Instruction Manual USER NOTES xiv 044100102 Rev A ...
Page 19: ...M200E Documentation Model 200E Instruction Manual USER NOTES 4 044100102 Rev A ...
Page 141: ...Calibration Procedures Model 200E Instruction Manual USER NOTES 126 044100102 Rev A ...
Page 203: ...Theory of Operation Model 200E Instruction Manual USER NOTES 188 044100102 Rev A ...
Page 239: ...Troubleshooting Repair Model 200E Instruction Manual USER NOTES 224 044100102 Rev A ...
Page 265: ...APPENDIX A Model 200E Instruction Manual USER NOTES 250 044100110 Rev A ...
Page 271: ...APPENDIX B Model 200E Instruction Manual USER NOTES 256 044100120 Rev A ...
Page 275: ...APPENDIX D Model 200E Instruction Manual 260 044100141 Rev A USER NOTES ...