
Model 200E Instruction Manual
Calibration Procedures
SAMPLE RANGE = 500.0 PPB NOX=X.XX
< TST TST > CAL
CALZ
CALS
SETUP
ZERO CAL M
RANGE = 500.0 PPB X=X.XX
< TST TST >
ZERO
CONC
EXIT
SAMPLE
RANGE TO CAL: LOW
LOW
HIGH
ENTR SETUP
SAMPLE
RANGE TO CAL: HIGH
LOW HIGH
ENTR
SETUP
…continue Calibration as per
standard procedure
Analyzer enters
ZERO
CAL
mode
Once this selection is made, the calibration/check procedure continues as described in
Sections 7.2 through 7.6. The other range may be calibrated by starting over from the main
SAMPLE
display.
7.7.2. Independent Range Mode
Independent range mode is described in Section 6.5.4. When selecting span gas concentra-
tions for the NO/NO
X
or NO
2
dynamic calibrations, use concentrations around 80% of the
highest reporting range for all gases. This will produce the most accurate calibration.
The span calibration routine is only required to be performed once, unlike AUTO range
mode, which requires separate
HIGH
and
LOW
range calibrations. If the calibration data
are obtained from the RS-232 port or from the front panel display, no special changes are
necessary as the
IND
range mode only affects the scaling of the analog outputs.
7.7.3. Calibration With Remote Contact Closures
Contact closures for controlling calibration and calibration checks are located on the rear
panel
CONTROL IN
connector. Instructions for setup and use of these contacts can be
found in Section 6.8.2.
When the appropriate contacts are closed for at least 5 seconds, the instrument switches
into zero, low span or high span mode and internal zero/span valves (if installed) will be
automatically switched to the appropriate configuration. The remote calibration contact
closures may be activated in any order. It is recommended that contact closures remain
closed for at least 10 minutes to establish a reliable reading; the instrument will stay in the
selected mode for as long as the contacts remain closed.
If contact closures are used in conjunction with the analyzer’s AutoCal (Section 7.8) feature
and the AutoCal attribute
CALIBRATE
is enabled, the M200E will not re-calibrate the
analyzer until the contact is opened. At this point, the new calibration values will be
recorded before the instrument returns to SAMPLE mode. If the AutoCal attribute
CALIBRATE
is disabled, the instrument will return to SAMPLE mode, leaving the instru-
ment’s internal calibration variables unchanged.
044100102 Rev A
121
Summary of Contents for 200E
Page 15: ...M200E Documentation Model 200E Instruction Manual USER NOTES xiv 044100102 Rev A ...
Page 19: ...M200E Documentation Model 200E Instruction Manual USER NOTES 4 044100102 Rev A ...
Page 141: ...Calibration Procedures Model 200E Instruction Manual USER NOTES 126 044100102 Rev A ...
Page 203: ...Theory of Operation Model 200E Instruction Manual USER NOTES 188 044100102 Rev A ...
Page 239: ...Troubleshooting Repair Model 200E Instruction Manual USER NOTES 224 044100102 Rev A ...
Page 265: ...APPENDIX A Model 200E Instruction Manual USER NOTES 250 044100110 Rev A ...
Page 271: ...APPENDIX B Model 200E Instruction Manual USER NOTES 256 044100120 Rev A ...
Page 275: ...APPENDIX D Model 200E Instruction Manual 260 044100141 Rev A USER NOTES ...