Model 200E Instruction Manual
Troubleshooting & Repair
11.4.1. Excessive noise
Excessive noise levels under normal operation usually indicate leaks in the sample supply or
the analyzer itself. Make sure that the sample or span gas supply is leak-free and carry out
a detailed leak check as described earlier in this chapter.
Another possibility of excessive signal noise may be the preamplifier board, the high voltage
power supply and/or the PMT detector itself. Contact the factory on trouble-shooting these
components.
11.4.2. Slow Response
If the analyzer starts responding too slow to any changes in sample, zero or span gas,
check for the following:
•
Dirty or plugged sample filter or sample lines.
•
Sample inlet line is too long.
•
Leaking NO/NO
X
valve. Carry out a leak check.
•
Dirty or plugged critical flow orifices. Check flows, pressures and, if necessary,
change orifices (Section 9.3.10).
•
Wrong materials in contact with sample - use glass, stainless steel or Teflon
materials only. Porous materials, in particular, will cause memory effects and slow
changes in response.
•
Dirty reaction cell. Clean the reaction cell.
•
Insufficient time allowed for purging of lines upstream of the analyzer. Wait until
stability is low.
•
Insufficient time allowed for NO or NO
2
calibration gas source to become stable. Wait
until stability is low.
•
NO
2
converter temperature is too low. Check for proper temperature.
11.4.3. Auto-zero Warnings
Auto-zero warnings occur if the signal measured during an auto-zero cycle is lower than –
20 mV or higher than 200 mV. The Auto-Zero warning displays the value of the auto-zero
reading when the warning occurs.
•
If this value is higher than 150 mV, check that the auto-zero valve is operating
properly. To do so, use the
SIGNAL I/O
functions in the
DIAG
menu to toggle the
valve on and off. Listen if the valve is switching, see if the respective LED on the
relay board is indicating functionality. Scroll the
TST
functions until
PMT
is displayed
and observe the PMT value change between the two valve states.
•
If the valve is operating properly, you should be able to hear it switch (once a
minute under normal operation or when manually activated from the
SIGNAL I/O
menu), the PMT value should drop from span gas reading (e.g., 800-900 mV at 400
ppb NO) to less than 150 mV and the LED on the relay board should light up when
the valve is activated. If the PMT value drops significantly but not to less than 150
mV, the valve is probably leaking across its ports. In this case, replace the valve. If
the PMT value does not change at all, the valve is probably not switching at all.
044100102 Rev A
203
Summary of Contents for 200E
Page 15: ...M200E Documentation Model 200E Instruction Manual USER NOTES xiv 044100102 Rev A ...
Page 19: ...M200E Documentation Model 200E Instruction Manual USER NOTES 4 044100102 Rev A ...
Page 141: ...Calibration Procedures Model 200E Instruction Manual USER NOTES 126 044100102 Rev A ...
Page 203: ...Theory of Operation Model 200E Instruction Manual USER NOTES 188 044100102 Rev A ...
Page 239: ...Troubleshooting Repair Model 200E Instruction Manual USER NOTES 224 044100102 Rev A ...
Page 265: ...APPENDIX A Model 200E Instruction Manual USER NOTES 250 044100110 Rev A ...
Page 271: ...APPENDIX B Model 200E Instruction Manual USER NOTES 256 044100120 Rev A ...
Page 275: ...APPENDIX D Model 200E Instruction Manual 260 044100141 Rev A USER NOTES ...