Operating Instructions
Model 200E Instruction Manual
Table 6-12: Current Loop Output Calibration with Resistor
Full scale
Voltage for 2-20 mA
(measured across resistor)
Voltage for 4-20 mA
(measured across resistor)
0%
0.5 V
1.0 V
100%
5.0 V
5.0 V
6.7.3.6. AIN Calibration
This is the sub-menu to conduct the analog input calibration. This calibration should only be
necessary after major repair such as a replacement of CPU, motherboard or power supplies.
Exit at any time to
return to the main
DIAG
menu
DIAG
ANALOG I / O CONFIGURATION
PREV
NEXT
ENTR
EXIT
DIAG AIO
AIN A/C FREQUENCY: 60 HZ
SET>
EDIT
EXIT
DIAG AIO
AIN CALIBRATED:
NO
< SET
SET>
CAL
EXIT
DIAG AIO
AIN CALIBRATED:
YES
< SET
SET> CAL
EXIT
Instrument
calibrates
automatically
Exit to return to the
ANALOG I/O
CONFIGURATION
MENU
STARTING FROM ANALOG I / O CONFIGURATION MENU
DIAG AIO
CALIBRATING A/D ZERO
CALIBRATING A/D SPAN
6.7.4. Test Channel Output
When activated, output channel
A4
can be used to report one of the test functions viewable
from the SAMPLE mode display.
68
044100102 Rev A
Summary of Contents for 200E
Page 15: ...M200E Documentation Model 200E Instruction Manual USER NOTES xiv 044100102 Rev A ...
Page 19: ...M200E Documentation Model 200E Instruction Manual USER NOTES 4 044100102 Rev A ...
Page 141: ...Calibration Procedures Model 200E Instruction Manual USER NOTES 126 044100102 Rev A ...
Page 203: ...Theory of Operation Model 200E Instruction Manual USER NOTES 188 044100102 Rev A ...
Page 239: ...Troubleshooting Repair Model 200E Instruction Manual USER NOTES 224 044100102 Rev A ...
Page 265: ...APPENDIX A Model 200E Instruction Manual USER NOTES 250 044100110 Rev A ...
Page 271: ...APPENDIX B Model 200E Instruction Manual USER NOTES 256 044100120 Rev A ...
Page 275: ...APPENDIX D Model 200E Instruction Manual 260 044100141 Rev A USER NOTES ...