Model 200E Instruction Manual
Theory of Operation
stable readings and averages out a considerable amount of random noise for an overall less
noisy concentration reading.
If the filter detects rapid changes in concentration the filter reduces the averaging to only 6
samples or about 48 seconds to allow the analyzer to respond more quickly. Two conditions
must be simultaneously met to switch to the short filter. First, the instantaneous concentra-
tion must differ from the average in the long filter by at least 50 ppb. Second, the instanta-
neous concentration must differ from the average in the long filter by at least 10% of the
average in the long filter.
Figure 10-18:
Schematic of Basic Software Operation
If necessary, these boxcar filter lengths can be
changed between 1 (no averaging) and
1000 samples but with corresponding tradeoffs in rise time and signal-to-noise ratio. Signal
noise increases accordingly when in adaptive filter mode, but remains within the official
M200E specifications defined by the EPA as long as the filter size remains at or above 3
samples.
In order to avoid frequent switching between the two filter sizes, the analyzer has a delay
of 120 s before switching out of adaptive filter mode, even if the two threshold conditions
are no longer met.
10.4.2. Calibration - Slope and Offset
Aside from the hardware calibration of the preamplifier board (Section 11.6.5) upon factory
checkout, calibration of the analyzer is usually performed in software. During instrument
calibration (Chapters 7 and 8) the user enters expected values for span gas concentration
through the front panel keypad and supplies the instrument with sample gas of know NO
and NO
X
concentrations. The readings are then compared to the expected values and the
software computes values for the new instrument slope and offset for both NO and NO
X
response. These values are stored in memory for use in calculating the NO, NO
X
and NO
2
concentration of the sample gas. By default, the iDAS stores 200 software calibration
settings for documentation, review and data analysis.
044100102 Rev A
185
Summary of Contents for 200E
Page 15: ...M200E Documentation Model 200E Instruction Manual USER NOTES xiv 044100102 Rev A ...
Page 19: ...M200E Documentation Model 200E Instruction Manual USER NOTES 4 044100102 Rev A ...
Page 141: ...Calibration Procedures Model 200E Instruction Manual USER NOTES 126 044100102 Rev A ...
Page 203: ...Theory of Operation Model 200E Instruction Manual USER NOTES 188 044100102 Rev A ...
Page 239: ...Troubleshooting Repair Model 200E Instruction Manual USER NOTES 224 044100102 Rev A ...
Page 265: ...APPENDIX A Model 200E Instruction Manual USER NOTES 250 044100110 Rev A ...
Page 271: ...APPENDIX B Model 200E Instruction Manual USER NOTES 256 044100120 Rev A ...
Page 275: ...APPENDIX D Model 200E Instruction Manual 260 044100141 Rev A USER NOTES ...