Model 200E Instruction Manual
Troubleshooting & Repair
•
Disconnect the heater cable coming from the relay board and measure the
resistance between any two of the three heater leads with a multi-meter. The
resistance between A and B should be about 1000
Ω
and that between A and C
should be the same as between B and C, about 500
Ω
each. If any of these
resistances is near zero or without continuity, the heater is broken.
•
Temperature reports zero or overload (near 500° C). This indicates a disconnected
or failing thermocouple or a failure of the thermocouple circuit.
•
First, check that the thermocouple is connected properly and the wire does not
show signs of a broken or kinked pathway. If it appears to be properly
connected, disconnect the yellow thermocouple plug (marked
K
) from the relay
board and measure the voltage (not resistance) between the two leads with a
multi-meter capable of measuring in the low mV range. The voltage should be
about 12 mV (ignore the sign) at 315° C and about 0 mV at room temperature.
•
Measure the continuity with an Ohm-meter. It should read close to zero
Ω
. If the
thermocouple does not have continuity, it is broken. If it reads zero voltage at
elevated temperatures, it is broken. To test the thermocouple at room tempera-
ture, heat up the converter can (e.g., with a heat gun) and see if the voltage
across the thermocouple leads changes. If the thermocouple is working properly,
the electronic circuit is broken. In both cases, consult the factory.
2) If the converter appears to have performance problems (conversion efficiency is outside
of allowed range of 96-102%), check the following:
•
Conversion efficiency setting in the CAL menu. If this value is different from 1.000,
this correction needs to be considered. Section 7.1.7 describes this parameter in
detail.
•
Accuracy of NO
2
source (GPT or gas tank standard). NO
2
gas standards are typically
certified to only ±2% and often change in concentrations over time. You should get
the standard re-certified every year. If you use GPT, check the accuracy of the ozone
source.
•
Age of the converter. The NO
2
converter has a limited operating life and may need to
be replaced every ~3 years or when necessary (e.g., earlier if used with continu-
ously high NO
2
concentrations). We estimate a lifetime of about 10000 ppm-hours (a
cumulative product of the NO
2
concentration times the exposure time to that
concentration). However, this lifetime heavily depends on many factors such as
absolute concentration (temporary or permanent poisoning of the converter is
possible), sample flow rate and pressure inside the converter, converter tempera-
ture, duty cycle etc. This lifetime is only an estimated reference and not a
guaranteed lifetime.
•
In some cases with excessive sample moisture, the oxidized molybdenum metal
chips inside the converter cartridge may bake together over time and restrict air flow
through the converter, in which case it needs to be replaced. To avoid this problem,
we recommend the use of a sample gas conditioner (Section 5.8). Section 9.3.8
describes how to replace the NO
2
converter cartridge.
•
With no NO
2
in the sample gas and a properly calibrated analyzer, the NO reading is
negative, while the NO
2
reading remains around zero. The converter destroys NO
and needs to be replaced.
•
With no NO
2
in the sample gas and a properly calibrated analyzer, the NO
X
reading is
significantly higher than the actual (gas standard) NO concentration. The converter
produces NO
2
and needs to be replaced.
044100102 Rev A
215
Summary of Contents for 200E
Page 15: ...M200E Documentation Model 200E Instruction Manual USER NOTES xiv 044100102 Rev A ...
Page 19: ...M200E Documentation Model 200E Instruction Manual USER NOTES 4 044100102 Rev A ...
Page 141: ...Calibration Procedures Model 200E Instruction Manual USER NOTES 126 044100102 Rev A ...
Page 203: ...Theory of Operation Model 200E Instruction Manual USER NOTES 188 044100102 Rev A ...
Page 239: ...Troubleshooting Repair Model 200E Instruction Manual USER NOTES 224 044100102 Rev A ...
Page 265: ...APPENDIX A Model 200E Instruction Manual USER NOTES 250 044100110 Rev A ...
Page 271: ...APPENDIX B Model 200E Instruction Manual USER NOTES 256 044100120 Rev A ...
Page 275: ...APPENDIX D Model 200E Instruction Manual 260 044100141 Rev A USER NOTES ...