5.8. Sample Gas Conditioners (Options 86 & 88)
Several permeation devices using Nafion
®
permeation gas exchange tubes are available for
applications with high moisture and/or moderate levels of NH
3
in the sample gas. This type
of sample conditioner is part of the standard M200E equipment to remove H
2
O and NH
3
from the ozone generator supply gas stream but can be purchased for the sample gas
stream as well. All gas conditioners remove water vapor to a dew point of about –20° C
(~600 ppm H
2
O) and effectively remove concentrations of ammonia up to about 1 ppm.
More information about these dryers and their performance is available at
http://www.permapure.com/.
The following options include the hardware required to install the dryers.
Table 5-4: Dryer and NH
3
Removal Options
Option Number
Description
Standard Equipment
Single gas conditioner (dryer / NH
3
removal) for ozone generator supply gas
stream only. Includes mounting bracket for two dryers (Option 86 mounts on
the back).
OPT 86
Single gas conditioner (dryer / NH
3
removal) for sample gas stream only.
Mounts on the back of the existing dryer bracket. Converts analyzer to dual-
conditioner instrument.
OPT 88
Single combination gas conditioner (dryer / NH
3
removal) for both the
sample gas and ozone supply air. Replaces the standard dryer for O
3
air and
comes with mounting bracket.
The combination conditioner is a low-cost option for drying both the sample gas and ozone
supply air with one dryer. However, this dryer can only be used in applications where both
sample and calibration gases (after dilution) are at or near ambient and constant concen-
trations of oxygen (about 20%), because the ozone generator needs a high and constant
amount of oxygen to generate ozone properly. Stack applications or industrial applications
044100102 Rev A
39
Summary of Contents for 200E
Page 15: ...M200E Documentation Model 200E Instruction Manual USER NOTES xiv 044100102 Rev A ...
Page 19: ...M200E Documentation Model 200E Instruction Manual USER NOTES 4 044100102 Rev A ...
Page 141: ...Calibration Procedures Model 200E Instruction Manual USER NOTES 126 044100102 Rev A ...
Page 203: ...Theory of Operation Model 200E Instruction Manual USER NOTES 188 044100102 Rev A ...
Page 239: ...Troubleshooting Repair Model 200E Instruction Manual USER NOTES 224 044100102 Rev A ...
Page 265: ...APPENDIX A Model 200E Instruction Manual USER NOTES 250 044100110 Rev A ...
Page 271: ...APPENDIX B Model 200E Instruction Manual USER NOTES 256 044100120 Rev A ...
Page 275: ...APPENDIX D Model 200E Instruction Manual 260 044100141 Rev A USER NOTES ...