![TEL AFA4000/E Installation And Operating Manual Download Page 77](http://html1.mh-extra.com/html/tel/afa4000-e/afa4000-e_installation-and-operating-manual_1079270077.webp)
AFA4000/E
VAV Airflow Controller / p.77
Figure 40: Mounting transmitter / receiver
6.3.3
Aligning the light curtain transmitter and receiver
Mount the light curtain receiver and transmitter at the same height on opposing sides of the
fume hood opening. A small error in alignment is acceptable. The allowable error is dependent
on the spacing between the transmitter and receiver.
Figure 41 illustrates the maximum allowable transmitter to receiver misalignment /
displacement for a 0 to 3 m (0 to 118.11") range of fume hood widths. For a typical fume hood
width of 1 to 2 m (39.37 to 78.740"), the maximum allowable misalignment / displacement
ranges from approximately 20 to 40 mm (0.79 to 1.58").
Summary of Contents for AFA4000/E
Page 1: ...AFA4000 E VAV Airflow Controller Installation and operating manual Issue 2...
Page 57: ...AFA4000 E VAV Airflow Controller p 57 Figure 22 AFA4000 E inverter control connections...
Page 58: ...AFA4000 E VAV Airflow Controller p 58 Figure 23 AFA4000 E damper control connections...
Page 59: ...AFA4000 E VAV Airflow Controller p 59 Figure 24 AFA4000 E typical connection diagram...
Page 60: ...AFA4000 E VAV Airflow Controller p 60 Figure 25 AFA4000 E AS typical connection diagram...
Page 62: ...AFA4000 E VAV Airflow Controller p 62 Figure 27 SM7 airflow sensor dimensions...
Page 96: ...AFA4000 E VAV Airflow Controller p 96 6 9 Auto Sash System Control Box general wiring diagram...
Page 158: ...AFA4000 E VAV Airflow Controller p 158...
Page 182: ...AFA4000 E VAV Airflow Controller p 182 Figure 79 Connection diagram with RS232 485 converter...