AFA4000/E
VAV Airflow Controller / p.31
Function
Description
Mains Fail
(Optional extra feature)
When the input configured as Mains Fail is activated:
•
Red Alarm LED Halo illuminated
•
Mains Fail
displays
•
Audible alarm sounds
•
Audible can be muted by pressing the
Mute
pushbutton. This
silences the alarm (when configured)
Note: Not for use with the Relay Interface Unit.
Personnel Sensor
(Optional extra feature)
When the input configured as Personnel Sensor is activated:
•
VAV Dual Set Point is active:
Occupied = High Set Point;
Unoccupied = Low Set Point
Note: When the Low Set Point is active, the Reduced Low Air and
Warning Air Alarms are active.
Fan Stop
When the input configured as Fan Stop is activated:
•
Off
displays
•
Red LED Halo illuminated
•
Audible alarm muted
When the input configured as Fan Stop is de-activated:
•
Start Up Timer
displays
•
Audible alarm and relays inhibited until Start up time has elapsed
Note: Used if the AFA4000 is a Slave unit with a common fan.
Remote master On/Off signal is used
Mute
When the input configured as Mute is activated:
•
Audible alarm muted
•
Mute Icon shown on display
Note: The Mute function disables the
Mute
pushbutton so that any
audible alarm can only be muted using the input, for example
remote key switch
Summary of Contents for AFA4000/E
Page 1: ...AFA4000 E VAV Airflow Controller Installation and operating manual Issue 2...
Page 57: ...AFA4000 E VAV Airflow Controller p 57 Figure 22 AFA4000 E inverter control connections...
Page 58: ...AFA4000 E VAV Airflow Controller p 58 Figure 23 AFA4000 E damper control connections...
Page 59: ...AFA4000 E VAV Airflow Controller p 59 Figure 24 AFA4000 E typical connection diagram...
Page 60: ...AFA4000 E VAV Airflow Controller p 60 Figure 25 AFA4000 E AS typical connection diagram...
Page 62: ...AFA4000 E VAV Airflow Controller p 62 Figure 27 SM7 airflow sensor dimensions...
Page 96: ...AFA4000 E VAV Airflow Controller p 96 6 9 Auto Sash System Control Box general wiring diagram...
Page 158: ...AFA4000 E VAV Airflow Controller p 158...
Page 182: ...AFA4000 E VAV Airflow Controller p 182 Figure 79 Connection diagram with RS232 485 converter...