AFA4000/E
VAV Airflow Controller / p.174
6.
Select
Sash Warning Timer
>
Sash Warning Time Delay.
7.
Use the
↑
/
↓
buttons to set the
Sash Warn Rpt Time
to
Enabled / Disabled
. If set to
Enabled
,
use the
↑
/
↓
buttons to set the
Sash Warn Rpt Time
, then
Done
.
For example: Sash closed switch height set to 100 mm, sash warning delay set to 1 minute,
sash warning repeat timer set to 5 minutes. If the sash is > 100 mm, the close sash alarm
activates if the fume cupboard is unoccupied for 1 minute. After being muted, the alarm re-
sounds after 5 minutes. The alarm resets if the fume cupboard is then occupied or the sash is
<100 mm.