AFA4000/E
VAV Airflow Controller / p.195
When a BMS input is in an active condition, the current alarm displays in the status window:
Alarm message
Description
Fire Alarm
This is displayed along with the normal text, for example
Fire
Alarm
or
User present,
when the fire alarm input is activated. The
sash will drive closed when the user is not detected and the sash
is clear. The sash can be operated manually when the user is
detected (auto open, tiptronic & pushbuttons are disabled).
BMS open
This is displayed along with the normal text when the user is not
detected and when the BMS Open input is activated. The sash will
drive open when the user is not present and the sash is clear. The
sash can be operated normally when the user is detected (auto
open, tiptronic & pushbuttons are active).
BMS close
This is displayed when the BMS Close input is activated, the sash
will drive closed when the user is not present and the sash is
clear. The sash can be operated normally when the user is
detected (auto open, tiptronic & pushbuttons are active).
Emergency (up)
This is displayed along with the normal text when the user is not
detected and when the EV Open input is activated. The sash will
drive open when the user is not present and the sash is clear. The
sash can be operated manually when the user is detected (auto
open, tiptronic & pushbuttons are disabled).
Emergency (dn)
This is displayed along with the normal text when the user is not
detected and when the EV Close input is activated. The sash will
drive closed when the user is not present and the sash is clear.
The sash can be operated manually when the user is detected
(auto open, tiptronic & pushbuttons are disabled).
Sash Opening
This is displayed when the Sash Open Footswitch has been
operated whilst the operator is present.
Sash Closing
This is displayed when the Sash Close Footswitch has been
operated whilst the operator is present.
Summary of Contents for AFA4000/E
Page 1: ...AFA4000 E VAV Airflow Controller Installation and operating manual Issue 2...
Page 57: ...AFA4000 E VAV Airflow Controller p 57 Figure 22 AFA4000 E inverter control connections...
Page 58: ...AFA4000 E VAV Airflow Controller p 58 Figure 23 AFA4000 E damper control connections...
Page 59: ...AFA4000 E VAV Airflow Controller p 59 Figure 24 AFA4000 E typical connection diagram...
Page 60: ...AFA4000 E VAV Airflow Controller p 60 Figure 25 AFA4000 E AS typical connection diagram...
Page 62: ...AFA4000 E VAV Airflow Controller p 62 Figure 27 SM7 airflow sensor dimensions...
Page 96: ...AFA4000 E VAV Airflow Controller p 96 6 9 Auto Sash System Control Box general wiring diagram...
Page 158: ...AFA4000 E VAV Airflow Controller p 158...
Page 182: ...AFA4000 E VAV Airflow Controller p 182 Figure 79 Connection diagram with RS232 485 converter...