AFA4000/E
VAV Airflow Controller / p.105
7.4
Over Current Monitor
1.
From the
Run Screen
press and hold the
Mute
button for 5 seconds until the
Main Menu
is
displayed.
2.
Use the
↑
/
↓
buttons to select
Set Up Auto Sash
. The display shows the message
TEL SASH
CLOSER
.
3.
Press the
Mute
button to display the current Hardware and Software version, for example,
1.
HW2.0 FW1.1
.
4.
Press the
↓
button to scroll down and select
Setup Menu
and press
Mute
.
5.
Enter the password (the factory default password is 1-0-0-1) and press
Mute
.
6.
In the
Configuration
menu the controller will display the message
System Profile
.
7.
Use the
↓
button to scroll through the menu and select
Sash Settings
, then press
Mute
.
8.
In the
Configuration menu
the display will show
1. Enabled
.
9.
Use the
↓
button to scroll through the menu and select
3. Motor Drive
. Press
Mute
.
10.
Use the
↓
button to scroll through the menu and select
Motor Current Display,
Press
Mute
.
11.
Use the
↑
/
↓
buttons to select
Max,
Press
Mute
.
Note
: Select Max to show the Peak current or Live to show the live value.
12.
Use the
↓
button and select
Back
then
Back to Set up menu
. Press
Mute
.
13.
Use the
↓
button and select
Exit and Save Changes
. Press
Mute
.
14.
When the Motor drives, the selected
Max
or
Live
current will be displayed in the status bar.
15.
Determine the current limit required for either sash open, sash close or both.
16.
From the
Run Screen
press and hold the
Mute
button for 5 seconds until the
Main Menu
is
displayed.
17.
Use the
↑
/
↓
buttons to select
Set Up Auto Sash
. The display shows the message
TEL SASH
CLOSER
.
18.
Press the
↓
button to scroll down and select
Setup Menu
and press
Mute
.
19.
Enter the password (the factory default password is 1-0-0-1) and press
Mute
.
20.
Use the
↓
button to scroll through the menu and select
Sash Settings
, then press
Mute
.
21.
Use the
↓
button to scroll through the menu and select
3. Motor Drive
. Press
Mute
.
22.
Use the
↓
button to scroll through the menu and select
Motor Current Display,
Press
Mute
.
23.
Use the
↑
/
↓
buttons to select
Off,
Press
Mute
.
24.
Use the
↓
button to scroll through the menu and select
Over current Open,
Press
Mute
.
25.
Use the
↑
/
↓
buttons to set the required limit
,
Press
Mute
.
26.
Use the
↓
button to scroll through the menu and select
Over current Close,
Press
Mute
.
27.
Use the
↑
/
↓
buttons to set the required limit
,
Press
Mute
Summary of Contents for AFA4000/E
Page 1: ...AFA4000 E VAV Airflow Controller Installation and operating manual Issue 2...
Page 57: ...AFA4000 E VAV Airflow Controller p 57 Figure 22 AFA4000 E inverter control connections...
Page 58: ...AFA4000 E VAV Airflow Controller p 58 Figure 23 AFA4000 E damper control connections...
Page 59: ...AFA4000 E VAV Airflow Controller p 59 Figure 24 AFA4000 E typical connection diagram...
Page 60: ...AFA4000 E VAV Airflow Controller p 60 Figure 25 AFA4000 E AS typical connection diagram...
Page 62: ...AFA4000 E VAV Airflow Controller p 62 Figure 27 SM7 airflow sensor dimensions...
Page 96: ...AFA4000 E VAV Airflow Controller p 96 6 9 Auto Sash System Control Box general wiring diagram...
Page 158: ...AFA4000 E VAV Airflow Controller p 158...
Page 182: ...AFA4000 E VAV Airflow Controller p 182 Figure 79 Connection diagram with RS232 485 converter...