Performance
Check—
DM 5010
Table
5-5
DC
NORMAL MODE REJECTION
50-60
Hz
Jumper
Position
DM
5010
FAST
Calibrator
Frequency
(Hz)
Display
Reading
Off
60.2, ±0.02
PASS
on
60.2 ±0.02
PASS
60 Hz
on
59.8, ±0.02
PASS
off
59.8, ±0.02
PASS
off
50.2, ±0.02
PASS
off
49.8, ±0.02
PASS
50 Hz
on
49.8, ±0.02
PASS
on
50.2, ±0.02
PASS
5.
Ohms
Accuracy
a. Set the
DM 5010 front-panel controls to the prelimi
nary control settings with the following exception:
OHMS
on
b. Connect a shorting plug between the DM 5010 HIGH
and
LOW input
connectors.
c. CHECK—that the display reads within the limits listed
in Table 5-6, at each conversion rate.
d. Set the
DM 5010
front panel controls to the 200 Ω
range and normal conversion
rate.
Press the NULL button.
Remove
the shorting plug and connect the decade resis
tance
box
to
the DM 5010 input connectors using
a coaxial
cable with
less
than 0.3 Ω. See Fig. 5-3A.
e. Set the decade
box resistance to the value listed in
Table
5-7.
f. CHECK—that the display reads within the limits
listed
in
the table.
g. Set the
DM 5010 CONVERSION RATE as shown in
the
next
line of
the table.
h. CHECK—that the display reads within the limits listed
in
the table.
i.
Set the
DM5010
range, CONVERSION RATE, and
the
decade
box resistance as listed in the next line of the
table.
j. Repeat
parts f through
i for the remaining lines of the
table,
except
change the equipment setup as shown in
Fig.
5-3B for the 20 MΩ range checks.
k.
Remove ail
connections to
the DM 5010.
Table 5-6
OHMS
OFFSET ACCURACY
DM
5010
Range
DM 5010
FAST
Display Limits
Ambient
Temp. Range
+
18 to
+ 28°C
Ambient
Temp.
Range
Oto
+ 18’C,
+28 to +50°C
200
Ω
Off
on
±
0.03 Ω
a
±0.1
Ωa
±0.12
Ω
a
±0.2Ω
a
2 kΩ
on
off
±
.001
kΩ
± .0002 Ω6
±
.002 kΩ
± .0007
kΩc
20
kΩ
off
on
±0.002
kΩ
±0.01
kΩ
±0.007
kΩ
±0.02 kΩ
200 kΩ
on
off
±0.1
kΩ
±0.02 kΩ
±0.2 kΩ
±0.07 kΩ
2 MΩ
off
on
± .0002 MΩ
± .001
MΩ
± .0007 MΩ
±
.002 MΩ
20
MΩ
on
off
±0.01
MΩ
±0.001
MΩ
±0.01
MΩ
±0.002 MΩ
a
-
With NULL, if NULL is not used add
±0.2 Q.
b
- With NULL, if NULL is not
used add
±0.0002 kΩ.
REV
OCT
1984
5
:9
Summary of Contents for DM 5010
Page 14: ...DM 5010 2994 00 DM 5010 Programmable Digital Multimeter xii ADD JUL 1986...
Page 27: ...Operating Instructions DM 5010 2994 03 Fig 2 3 DM 5010 front panel controls and connectors 2 3...
Page 38: ......
Page 134: ......
Page 208: ......
Page 222: ......
Page 250: ......
Page 251: ...Section 8 DM 5010 OPTIONS No options are available 8 1...
Page 252: ......
Page 270: ......
Page 272: ...DM 5010 2994 37 Fig 10 2 Location of DM 5010 adjustments and test points...
Page 273: ......
Page 274: ......
Page 275: ......
Page 276: ...DM 5010 2994 112 DM 5010 BLOCK DIAGRAM...
Page 281: ......
Page 282: ......
Page 291: ......
Page 293: ......
Page 294: ......
Page 297: ......
Page 298: ......
Page 303: ......
Page 304: ...I...
Page 305: ......
Page 310: ......
Page 311: ......
Page 315: ......
Page 318: ......
Page 321: ......
Page 323: ......
Page 326: ......
Page 332: ...2994 57...
Page 334: ......
Page 335: ......
Page 336: ......
Page 337: ...63 REV JUN 1986...
Page 338: ...FIG 1 EXPLODED DM 5010...
Page 339: ......
Page 340: ......
Page 341: ......
Page 347: ......