Section
5—
DM 5010
PERFORMANCE
CHECK
Introduction
This performance
check verifies the Electrical Perfor
mance Requirements as listed in the Specification section in
this
manual. Perform the
Adjustment Procedure if the instru
ment
fails to meet these checks. For convenience, many
steps
in this procedure check the performance of this instru
ment
at only one value in the specified
performance range.
Any value within
the specified range, within appropriate lim
its, may
be substituted. The performance check
may
be
done
at any ambient temperature between 0°
and
+50°C.
Performance
limits for two ambient temperature ranges are
listed
for some
steps.
Use the data listed for the tempera
ture at which the instrument is operating. A
summary sheet
is
provided at the back
of this section for recording
perfor
mance check results.
Test
Equipment Required
The
test equipment listed
in Table 5-1, or equivalent,
is
recommended to perform the
performance
check. Specifica
tions given for the test equipment are the minimum neces
sary
for
accurate
performance verification. All
test
equipment is
assumed to be correctly calibrated and operat
ing within specification.
Table
5-1
TEST EQUIPMENT LIST
Description
Minimum
Requirements
Performance
Check Step
Recommended
Equipment
TM 5000-Series Power
Module
all
TEKTRONIX TM
5003,
Opt.
02
Dc voltage calibrator
±200
mV: ±(0.0038% of
rdng
+
0.0025% of range)
±2V-200V:
±(0.0038% of
rdng
+ 0.0013% of range)
±1000 V:
±(0.005% of rdng +
0.0025% of
range)
1. 2, 3
Fluke
335D
Dc Voltage Standard
Ac
voltage calibrator
200
mV
through 200 V:
10-20 Hz: ±(0.20% of rdng
+
0.08%
of
range)
20-100 Hz: ±(0.20% of rdng
+ 0.05% of
range)
100Hz-20kHz:
±(0.05%
of
rdng
+ 0.05% of range)
20-100
kHz:
± (0.25% of rdng +
0.12%
of
range)
2.
3.4
Fluke 5200A
Ac Calibrator,
and
Fluke
5215A Power
Amplifier
700 V:
10-20 Hz:
±(0.20% of rdng
+
0.22%
of
range)
20-100 Hz: ±(0.20% of
rdng
+
0.15%
of range)
100Hz-15kHz:
±(0.05% of
rdng
+ 0.15% of range)
5-1
Summary of Contents for DM 5010
Page 14: ...DM 5010 2994 00 DM 5010 Programmable Digital Multimeter xii ADD JUL 1986...
Page 27: ...Operating Instructions DM 5010 2994 03 Fig 2 3 DM 5010 front panel controls and connectors 2 3...
Page 38: ......
Page 134: ......
Page 208: ......
Page 222: ......
Page 250: ......
Page 251: ...Section 8 DM 5010 OPTIONS No options are available 8 1...
Page 252: ......
Page 270: ......
Page 272: ...DM 5010 2994 37 Fig 10 2 Location of DM 5010 adjustments and test points...
Page 273: ......
Page 274: ......
Page 275: ......
Page 276: ...DM 5010 2994 112 DM 5010 BLOCK DIAGRAM...
Page 281: ......
Page 282: ......
Page 291: ......
Page 293: ......
Page 294: ......
Page 297: ......
Page 298: ......
Page 303: ......
Page 304: ...I...
Page 305: ......
Page 310: ......
Page 311: ......
Page 315: ......
Page 318: ......
Page 321: ......
Page 323: ......
Page 326: ......
Page 332: ...2994 57...
Page 334: ......
Page 335: ......
Page 336: ......
Page 337: ...63 REV JUN 1986...
Page 338: ...FIG 1 EXPLODED DM 5010...
Page 339: ......
Page 340: ......
Page 341: ......
Page 347: ......