Maintenance
52
P6467 High-Speed Probe Adapter Instruction Manual
H
Do not slide the probe over any surface.
H
Avoid handling the probe in areas that have a floor or work surface cover
that is capable of generating a static charge.
CAUTION.
To prevent damage to the probe, always disconnect the probe from the
DAS or TLA mainframe and from the system-under-test before servicing or
cleaning the probe.
Inspection and Cleaning
The P6467 Probe is inspected mechanically and electrically before shipment
from the factory. The probe should be free of marks or scratches and should meet
and exceed all electrical specifications. To confirm this, inspect the probe for
physical damage during transit. Save the probe packaging in case shipment or
repair is necessary. Contact your local Tektronix representative if there is any
damage or deficiency.
Cleaning procedures consist of exterior cleaning only. Periodic cleaning reduces
instrument failure and increases reliability. Clean the probe as needed, based on
the operating environment.
CAUTION.
Do not use chemical cleaning agents; they may damage the probe.
Avoid using chemicals that contain benzene, toluene, xylene, acetone, or
similar solvents.
Clean the exterior surfaces of the probe with a dry lint-free cloth or a soft-bristle
brush. If any dirt remains, use a cloth or swab dipped in a 75% isopropyl alcohol
solution. Use a swab to clean around controls or connectors. Do not use abrasive
compound on any part of the probe.
Removal and Replacement Instructions
The P6467 Probe has a limited number of components that can be replaced. All
replaceable parts are listed in the Replaceable Parts chapter beginning on
page 55. You can access all of components on the probe after removing the top
cover. Remove the top cover by removing the two screws on each side of the
probe.
Summary of Contents for 070-9176-00
Page 5: ......
Page 13: ...Service Safety Summary viii P6467 High Speed Probe Adapter Instruction Manual ...
Page 39: ...Operating Basics 24 P6467 High Speed Probe Adapter Instruction Manual ...
Page 45: ...Reference 30 P6467 High Speed Probe Adapter Instruction Manual ...
Page 49: ...Specifications 34 P6467 High Speed Probe Adapter Instruction Manual ...
Page 51: ......
Page 65: ...Adjustment Procedures 50 P6467 High Speed Probe Adapter Instruction Manual ...
Page 77: ...Replacable Parts 62 P6467 High Speed Probe Adapter Instruction Manual ...
Page 81: ...Appendix B P6467 Test Board 66 P6467 High Speed Probe Adapter Instruction Manual ...
Page 84: ......