![Tektronix 070-9176-00 Instruction Manual Download Page 55](http://html1.mh-extra.com/html/tektronix/070-9176-00/070-9176-00_instruction-manual_1077403055.webp)
Performance Verification
40
P6467 High-Speed Probe Adapter Instruction Manual
f.
Delete Sections D3 and D2 from the Data group.
g. Delete Sections C3, C1, and C0 from the Control group.
h. Ensure that the Input Radix is set to Hex for the remaining groups.
9. Select the Clock menu and perform the following steps:
a. Change the Clock field to External.
b. Select the rising edge of Clock_2 with no qualification.
10. Select the Trigger menu change it to match Figure 14.
Figure 14: Trigger Menu Setup
11. Repeat steps 7 through 10 for the module_b setups with the following
exceptions:
a. In the Clock menu, select the falling edge of Clock_2 instead of the
rising edge.
b. In the Trigger menu, assert Signal BtoA instead of Signal AtoB, and
change the last event to monitor Signal AtoB instead of Signal BtoA.
Summary of Contents for 070-9176-00
Page 5: ......
Page 13: ...Service Safety Summary viii P6467 High Speed Probe Adapter Instruction Manual ...
Page 39: ...Operating Basics 24 P6467 High Speed Probe Adapter Instruction Manual ...
Page 45: ...Reference 30 P6467 High Speed Probe Adapter Instruction Manual ...
Page 49: ...Specifications 34 P6467 High Speed Probe Adapter Instruction Manual ...
Page 51: ......
Page 65: ...Adjustment Procedures 50 P6467 High Speed Probe Adapter Instruction Manual ...
Page 77: ...Replacable Parts 62 P6467 High Speed Probe Adapter Instruction Manual ...
Page 81: ...Appendix B P6467 Test Board 66 P6467 High Speed Probe Adapter Instruction Manual ...
Page 84: ......