Performance Verification
38
P6467 High-Speed Probe Adapter Instruction Manual
H
Oscilloscope with FET probes with bandwidth sufficient to measure relative
time delays of approximately 5 ns
H
50
Termination (Tektronix part number 011-0049-01)
H
One 42-inch 50
coaxial cable (Tektronix part number 012-0057-01)
Performance Verification Procedure
Before attempting any of the procedures in this section, allow a minimum of a 20
minutes to warm up all test equipment. These instructions assume that you
perform the performance verification check while abiding by the environmental
conditions as described in Table 7 on page 32.
The following procedures are designed to test one P6467 Probe at a time. Repeat
the procedures for each additional probe.
The following steps describe the test equipment setups for the logic analyzer and
for the HFS 9003 Stimulus System. These steps assume that your DAS or TLA
logic analyzer already have the 92A96 Modules installed and that the Probe
Cables are properly connected to the 92A96 Modules. These steps also assume
that you have installed the P6467 Application Support software on the logic
analyzer.
Logic Analyzer Setups.
The following steps describe how to connect the probes to
the logic analyzer and how to set up the logic analyzer menus for the perfor-
mance verification checks.
NOTE
. The following procedures assume that the 92A96 Modules are located in
slots two and three of the logic analyzer. If your logic analyzer has the 92A96
Modules in different slots, you must adjust these procedures accordingly.
1. Remove the cover from each P6467 Probe that you are testing and set the
cover aside.
2. Connect the probe power cable from one of the 92A96 Modules in the logic
analyzer to the power connector on the P6467 Probe (if necessary, refer to
Figure 5 on page 7 for connection information). Leave the unused end of the
probe power cable disconnected.
3. Connect the blue-labeled coaxial probes from the logic analyzer to the
OUTPUT DATA A Connector (top or J100) of the P6467 Probe.
4. Power on the logic analyzer and wait for the Menu Selection overlay to
display.
Test Equipment Setups
Summary of Contents for 070-9176-00
Page 5: ......
Page 13: ...Service Safety Summary viii P6467 High Speed Probe Adapter Instruction Manual ...
Page 39: ...Operating Basics 24 P6467 High Speed Probe Adapter Instruction Manual ...
Page 45: ...Reference 30 P6467 High Speed Probe Adapter Instruction Manual ...
Page 49: ...Specifications 34 P6467 High Speed Probe Adapter Instruction Manual ...
Page 51: ......
Page 65: ...Adjustment Procedures 50 P6467 High Speed Probe Adapter Instruction Manual ...
Page 77: ...Replacable Parts 62 P6467 High Speed Probe Adapter Instruction Manual ...
Page 81: ...Appendix B P6467 Test Board 66 P6467 High Speed Probe Adapter Instruction Manual ...
Page 84: ......