P6467 High-Speed Probe Adapter Instruction Manual
51
Maintenance
This chapter provides procedures and information for inspecting and cleaning the
P6467 High-Speed Probe Adapter, removing and replacing probe components,
and isolating problems to specific components.
Service Strategy
The basic service strategy is to isolate faults to a module level and replace any
faulty modules with good modules. Any modules requiring repair should be
returned to your local Tektronix service center.
Performance verification can be accomplished by following the procedures in the
Performance Verification chapter beginning on page 37. There are no diagnostics
available for the P6467 assemblies. However, power-on diagnostics exist for the
92A96 Data Acquisition Modules. Refer to the DAS 9200 Technician’s Reference
Manual for information on the power-on diagnostics for the 92A96 Modules.
Preparation
The information in this chapter is designed for use by qualified service person-
nel. Read the Safety Summary at the front of this manual and the Service Safety
Summary before attempting any procedures in this chapter. Refer to the
Operating Basics chapter for information on the location of jumpers and probe
connections.
CAUTION.
Components on the P6467 Probe can be damaged by static discharge.
Service the probe only in a static-free environment. Observe standard handling
precautions for static-sensitive devices while servicing the probe. Always wear a
grounded wrist strap, or equivalent, while servicing the probe.
Observe the following precautions to avoid damaging the probe:
H
Do not handle static-sensitive components on the probe.
H
Transport and store the probe in its original container or on conductive foam.
Label any package that contains static-sensitive assemblies.
H
Wear a wrist strap to discharge static electricity from your body.
H
Do not allow anything capable of generating or holding a static charge on the
work surface.
Summary of Contents for 070-9176-00
Page 5: ......
Page 13: ...Service Safety Summary viii P6467 High Speed Probe Adapter Instruction Manual ...
Page 39: ...Operating Basics 24 P6467 High Speed Probe Adapter Instruction Manual ...
Page 45: ...Reference 30 P6467 High Speed Probe Adapter Instruction Manual ...
Page 49: ...Specifications 34 P6467 High Speed Probe Adapter Instruction Manual ...
Page 51: ......
Page 65: ...Adjustment Procedures 50 P6467 High Speed Probe Adapter Instruction Manual ...
Page 77: ...Replacable Parts 62 P6467 High Speed Probe Adapter Instruction Manual ...
Page 81: ...Appendix B P6467 Test Board 66 P6467 High Speed Probe Adapter Instruction Manual ...
Page 84: ......