Measuring Waveforms Automatically
3–70
TLS 216 User Manual
Table 3–3: Measurement Definitions (Cont.)
Name
Definition
Positive Width
Timing measurement of the first pulse in the waveform or gated region. The distance (time) between
MidRef (default 50%) amplitude points of a positive pulse.
Rise time
Timing measurement. Time taken for the leading edge of the first pulse in the waveform or gated
region to rise from a Low Ref value (default = 10%) to a High Ref value (default = 90%) of its final
value.
RMS
Voltage measurement. The true Root Mean Square voltage over the entire waveform or gated
region.
Setup
Timing measurement. The time between the last transition to a legal state for all waveforms in a
group (except the clock) and the clocking edge of the waveform specified as the clock. The clock can
be any channel. It does not have to be included in the measured group. Logic levels are dictated by
High Ref and Low Ref.
Skew
Timing measurement. The time between the first transition to an illegal state for all waveforms in a
group and last transition to a legal state for the same waveforms. Logic levels are dictated by High
Ref and Low Ref.
To Measure a Single Waveform
To take an automatic measurement, you must do the following steps.
1. Obtain a stable display of the waveform to be measured. Pressing AUTO-
SET may help.
2. Select a group and channel. The article Selecting Groups and Channels on
page 3–117 explains how to do these tasks.
3. Access the Measure menu, and turn on a measurement.
Do the following procedures to take single channel automatic measurements.
To select your single channel measurement, do the following steps.
1. Use the group selection buttons to select a group. Then use the Group menu
to select a channel.
2. Press MEASURE
➞
Select Measrmnt (main). (See Figure 3-32.)
3. Select a measurement from the side menu.
When taking measurements, consider the following hints.
H
To determine the channel on which the logic scope takes the measurement,
check the measurement readout.
Select a Measurement
Summary of Contents for P6240
Page 5: ......
Page 7: ......
Page 15: ...Table of Contents viii TLS 216 User Manual...
Page 20: ...Getting Started...
Page 26: ...Start Up 1 6 TLS 216 User Manual...
Page 51: ...Operating Basics...
Page 53: ...Overview 2 2 TLS 216 User Manual...
Page 73: ...Acquisition 2 22 TLS 216 User Manual...
Page 82: ...Reference...
Page 94: ...Adjusting the Vertical Setup 3 12 TLS 216 User Manual...
Page 104: ...Choosing an Acquisition Mode 3 22 TLS 216 User Manual...
Page 112: ...Choosing a Group Display Mode 3 30 TLS 216 User Manual...
Page 122: ...Customizing the Display 3 40 TLS 216 User Manual...
Page 130: ...Customizing the Display Colors 3 48 TLS 216 User Manual...
Page 162: ...Measuring Waveforms Automatically 3 80 TLS 216 User Manual...
Page 168: ...Measuring Waveforms with Cursors 3 86 TLS 216 User Manual...
Page 178: ...Printing a Hardcopy 3 96 TLS 216 User Manual...
Page 186: ...Probing of Circuits 3 104 TLS 216 User Manual...
Page 198: ...Saving and Recalling Waveforms 3 116 TLS 216 User Manual...
Page 222: ...Triggering on Patterns 3 140 TLS 216 User Manual...
Page 252: ...Triggering on States 3 170 TLS 216 User Manual...
Page 256: ...Viewing Status 3 174 TLS 216 User Manual...
Page 261: ...Appendices...
Page 283: ...Appendix B Algorithms B 18 TLS 216 User Manual...
Page 285: ...Appendix C Packaging for Shipment C 2 TLS 216 User Manual...
Page 290: ...Glossary...
Page 302: ...Glossary Glossary 12 TLS 216 User Manual...
Page 303: ...Index...
Page 319: ...Index Index 16 TLS 216 User Manual...
Page 320: ......
Page 321: ......