USAGE INSTRUCTION
After having installed the autoclave, proceed with preparations and use.
8.1 Turn on the autoclave
Press the general switch
(PICT.2 pos..
19
.).
8.2 Loading pure water tank
Connect the water filling pipe (provided) with the connection for water filling pipe and
plastic funnel PICT.14. Then engage the connection in the charging hole (PICT.2 pos.
.
13
.).
Insert the other end of the pipe with filter in the demineralized or distilled water container.
Press the button
B-PUMP
to start the water loading pump and press it until the countdown
start.
The pump loads the tank inside the autoclave. If maximum level is not reached within 180
seconds the pump automatically stops, therefore it is necessary to press the button again.
The pump stops automatically when maximum level is reached.
With connection to a deionizer, if the maximum water level has not been reached, the
functioning of the autoclave will be obstructed.
8.3 Manual filling
If you wish to use this method of filling or in the event of malfunctioning of the pump, the
filling of the tank can be carried out manually by the operator in the following way:
-Turn on autoclave
-Remove the connection cover of PICT.2 pos..
12
.
-Insert in this opening the hose connected to the funnel (PICT.
14
.)
-Pour distilled water in the funnel, keeping it higher than the loading top
-Add until the led of maximum level turns on (PICT.2 pos.
29
.).
-Loading completed, twist on the top
Summary of Contents for Europa B EVO
Page 2: ......
Page 9: ...7 ITALIANO FIG 2 ...
Page 40: ...38 ITALIANO FIG 32 ...
Page 52: ...7 ENGLISH PICT 2 ...
Page 83: ...38 ENGLISH PICT 32 ...
Page 95: ...7 FRANÇAIS FIG 2 ...
Page 126: ...38 FRANÇAIS FIG 32 ...
Page 138: ...7 ESPANOL FIG 2 ...
Page 169: ...38 ESPANOL FIG 32 ...
Page 181: ...7 DEUTCH FIG 2 ...
Page 212: ...38 DEUTCH ABB 32 ...