8.4 Characteristics of the water to be used
TABLE SHOWING THE QUALITY LEVELS LAID BY THE UNI EN 13060: 2005
CEN STANDARD UNI EN 13060 : 2005
Evaporation residue
≤
10
mg/l
Silicon oxide
≤
1
mg/l
Iron
≤
0.2
mg/l
Cadmium
≤
0.005
mg/l
Lead
≤
0.05
mg/l
mg/l Remains of heavy metals apart
from iron, cardium, lead
≤
0.1
mg/l
Chloride (Cl’)
≤
2
mg/l
Phosphate (P20s)
≤
0.5
mg/l
Conductivity (at 20°C)
≤
15
µs/cm
Ph value (acidity level)
5
÷÷÷÷
7,5
Appearance
Transparent, clear, without
deposits
Hardness (and alkaline earth ions)
≤
0.02
mmol/l
8.5 Loading the material into the autoclave
Place the materials to be sterilized on the trays provided.
- never superpose the materials
-
arrange the wrapped instruments, with the paper part facing upwards
-
never bring the materials into contact with the sterilization chamber or the closing door
-
put the scissors and dental forceps with the blade open
PICT.23
When you have place all the instruments close the door and you can read DOOR
CLOSED on the display.
8.6 Starting the sterilization cycle
Choose sterilization program most adaptable of load prepared and press the button
B-SELECT
.
Each individual pressure changes program selection to the next.
After having chosen the program start the cycle by pressing the button
B-START
, the door
will lock automatically and the cycle will begin.
During the cycle, the display will show all parameters and relative information.
Summary of Contents for Europa B EVO
Page 2: ......
Page 9: ...7 ITALIANO FIG 2 ...
Page 40: ...38 ITALIANO FIG 32 ...
Page 52: ...7 ENGLISH PICT 2 ...
Page 83: ...38 ENGLISH PICT 32 ...
Page 95: ...7 FRANÇAIS FIG 2 ...
Page 126: ...38 FRANÇAIS FIG 32 ...
Page 138: ...7 ESPANOL FIG 2 ...
Page 169: ...38 ESPANOL FIG 32 ...
Page 181: ...7 DEUTCH FIG 2 ...
Page 212: ...38 DEUTCH ABB 32 ...