29
ENGLISH
A 801
Irregular unloading pressure
Reset and restart cycle
A 901
Main voltage too low - Temporary lack of
electric energy
Reset and restart cycle,
Inform eletric Energy
provider
REPLACE RESINS
Resins IONIC
SYSTEM exhaust
Replacement of resins; refer to paragraph
14.3
REPLACE FILTER
Resins OSMOSI
SYSTEM exhaust
Replacement of filter; refer to paragraph
14.3
D
The printer’s door is not
closed
Try to close the printer’s door
P
Paper is missing
Insert paper
Lo There is low voltage on
the line
Try to connect the machine to the net electrical worker in
an other point; if the problem persists to contact a
electrician
IMPORTANT
If after following the remedies, the problem persists, call telephone assistance at
MEDLINE ITALIA srl +39 0522 – 94 29 97.