9
ENGLISH
02
SAFETY
2.1
Safety Marking
HAZARDOUS VOLTAGE
DISCONNECT THE POWER BEFORE REMOVING THE
LID
HIGH TEMPERATURE
EARTH CONNECTION
2.2
Safety Devices
The following Safety Devices are installed:
- Safety valve set at 2.4 bar 0/+10%
- Electromagnetic lock to prevent the door from opening while the cycle is running
- Resistance over temperature thermostats
2.3
Safety Notes
- The manufacturer is liable for the marketed product in accordance with current
regulations.
The manufacturer's liability will expire when operations are carried out on the device, or a
part of it, by unskilled personnel or using non-original spare parts.
- There should be no potential risk of explosion and/or fire in the room where the
autoclave is installed.
- The autoclave should be installed in a special well-ventilated room.
The incubator’s connector (PICT.1 pos.
18
.), should be used only for the B-TEST.
The 230 V a.c. connection is indipendent by the
MAINS
, don’t touch the connector with
wet hands and protect by the water.