60
A-96.250.701 / 140716
AMI INSPECTOR Oxygen
Program List and Explanations
5.3.1.2
Sample Flow:
Define at which sample flow a flow alarm should be
issued.
5.3.1.2.1
Flow Alarm:
Program if the alarm relay should be activated if there
is a flow alarm. Choose between yes or no. The flow alarm will
always be indicated in the display, pending error list, saved in the
message list and the logger.
Available values: Yes or no
NOTICE:
Sufficient flow is essential for a correct measurement.
We recommend to program yes.
5.3.1.2.2
Alarm High:
If the measuring values rises above the programmed
value E009 will be issued.
Range: 12–50 l/h
5.3.1.2.32
Alarm Low:
If the measuring values falls below the programmed
value E010 will be issued.
Range: 8–11 l/h
5.3.1.3
Sample Temp.:
Define at which sample temperature an alarm
should be issued.
5.3.1.3.1
Alarm High:
If the measured value rises above the alarm high
value, the alarm relay is activated and E007 is issued.
Range: 30–100 °C
5.3.1.3.22
Alarm Low:
If the measured value rises above the alarm high value,
the alarm relay is activated and E008 is issued.
Range: -10 to + 20 °C
5.3.1.4
Alarm Saturation
5.3.1.4.1
Alarm High:
If the measured value rises above the alarm high
value, the alarm relay is activated and E001, is displayed in the
message list.
Range: 0.00 –200 %
5.3.1.4.22
Alarm Low:
If the measured value falls below the alarm low value,
the alarm relay is activated and E002 is displayed in the message
list.
Range: 0.00 –200 %
5.3.1.4.32
Hysteresis:
Within the hyst. range, the relay does not switch. This
prevents damage of relays contacts when the measured value
fluctuates around the alarm value.
Range. 0.00 –200 %
5.3.1.4.42
Delay:
Duration, the activation of the alarm relay is retarded after
the measuring value has risen above/fallen below the programmed
alarm.
Range: 0–28‘800 Sec