AMI INSPECTOR Oxygen
Program List and Explanations
A-96.250.701 / 140716
61
5.3.1.5
Case Temp.
5.3.1.5.1
Case Temp. high:
Set the alarm high value for temperature of
electronics housing. If the value rises above the programmed value
E013 is issued.
Range: 30–75 °C
5.3.1.5.2
Case Temp. low:
Set the alarm low value for temperature of
electronics housing. If the value falls below the programmed value
E014 is issued.
Range: -10 to + 20 °C
5.3.2 and 5.3.3
Relay 1 and 2:
The function of relay contacts 1 or 2 are defined by
the user
NOTICE:
The navigation in the menu <Relay 1> and <Relay 2>
is equal. For reason of simplicity only the menu numbers of
Relay 1 are used in the following.
1
First select the functions as:
- Limit upper/lower,
- Control upwards/downwards,
- Timer
- Fieldbus
2
Then enter the necessary data depending on the selected func-
tion. The same values may also be entered in menu
4.2 Relay
Contacts, p. 53
When the relays are used as upper or lower limit switches, program
the following:
5.3.2.20
Parameter:
choose one of the following process values
Oxygen
Temperature
Sample Flow
Saturation
5.3.2.300
Setpoint
: If the measured value rises above respectively falls below
the set-point, the relay is activated.
5.3.2.1
Function = Limit upper/lower:
Parameter
Range
Oxygen
0.00 ppb –20.00 ppm
Temperature
-30 to + 130 °C
Sample flow
0–50 l/h
Saturation
0–200 %