AMI INSPECTOR Oxygen
Error List
E013
Case Temp. high
– check case/environment temperature
– check programmed value,
see
5.3.1.5.1, p. 61
E014
Case Temp. low
– check case/environment temperature
– check programmed value,
see
5.3.1.5.2, p. 61
E017
Control Timeout
– check control device or programming in
Installation, Relay contact, Relay 1/2
see
5.3.2 and 5.3.3, p. 61
E019
Quality Assurance
– Perform QA Procedure using a
reference instrument, e.g. AMI Inspector
E024
Input active
– See If Fault Yes is programmed in Menu
see
5.3.4, p. 65
E026
IC LM75
– call service
E030
EEProm Frontend
– call service
E031
Calibration Recout
– call service
E032
Wrong Frontend
– call service
E033
Power-on
– none, normal status
E034
Power-down
– none, normal status
E065
Electrolyte depleted
– Refill electrolyte, see
Electrolyte
exchange, p. 29
Error
Description
Corrective action