AMI INSPECTOR Oxygen
Maintenance
A-96.250.701 / 140716
29
6.3.
Maintenance of the Oxygen Sensor
WARNING
Etching liquid
The electrolyte is alkaline and caustic. It contains less than 1%
of potassium hydroxide.
Do not ingest. Wear protective goggles and gloves during
handling. Avoid contact with clothes.
In case of accidental contact with the eyes, wash immediately
with clear water and contact a physician. Show him the label
of the bottle or this section of the manual.
Short contact with skin is harmless, nevertheless wash with
lots of water.
6.3.1
Electrolyte exchange
An electrolyte exchange is indicated in the maintenance list as
soon as the remaining amount is below 10%.
1
Unscrew the fixing sleeve [A].
2
Remove the sensor from the flow cell.
A
B
C
Fixing sleeve
Sensor
Flow cell
A
C
B