AMI INSPECTOR Oxygen
Maintenance
A-96.250.701 / 140716
33
A
Tilted sensor
B
Flow cell
7
Press [Enter] to start the calibra-
tion measurement.
The saturation should reach
100%, the saturation current
should be about 24
A to 41
A.
If the measuring values are not
stable during the measuring
period, the calibration will be
discarded.
8
Press [Enter] to confirm the cali-
bration.
3.1.5
Calibration
Place the electrode into
the wet flow cell at
a slightly tilted angle.
<Enter> to continue
A
B
3.1.1
Calibration
Saturation
98.7 %
Sat. Current
32
A
Progress
3.1.1
Calibration
Saturation
98.7 %
Sat. Current
32
A
Calibration Successful