AMI INSPECTOR Oxygen
Instrument Setup
A-96.250.701 / 140716
23
4.
Instrument Setup
4.1.
Establish sample flow
1
Open flow regulating valve, see
Fluidics, p. 9
.
2
Wait until the flow cell has been completely filled.
3
Switch on power.
4
Adjust the sample flow to 8–25 l/h.
5
Let the instrument run-in for 1 h.
4.2.
Programming
Programming
Program all parameters for external devices (interface, recorders,
etc.)
Program all parameters for instrument operation (limits, alarms),
see
Program List and Explanations, p. 50