AMI INSPECTOR Oxygen
Program List and Explanations
3 Maintenance
3.1 Calibration
3.1.1
Start a calibration and follow the instructions on the screen.
Displayed values are saturation in % and the saturation current in
mA. The indication bar shows the progress. Detailed explanation
see
Calibration, p. 32
.
3.2 Service
3.2.1
Electrolyte
Last Filling
: Shows the date of the last filling of electrolyte.
Remaining Amount
: Remaining amount of electrolyte in %.
Remaining Time
: Remaining time in days until electrolyte exchange
recommended.
3.2.1.5
New Filling
: Select “Yes” after electrolyte exchange to reset the
counter.
3.3 Simulation
To simulate a value or a relay state, select the
alarm relay
relay 1 and 2
signal output 3 (signal outputs 1 and 2 are deactivated)
with the [
] or [
] keys.
Press the [Enter]> key.
Change the value with the [
] or [
] keys.
Press the [Enter] key.
The value is simulated by the relay/signal output.
At the absence of any key activities, the instrument will switch back
to normal mode after 20 min. If you quit the menu, all simulated
values will be reset.
3.4 Set Time
Adjust date and time.
Alarm Relay:
Active or inactive
Relay 1:
Active or inactive
Relay 2:
Active or inactive
Signal Output 3:
Current in mA (if option is installed)