AMI INSPECTOR Oxygen
Installation
A-96.250.701 / 140716
15
3.2.2
Connect the Sample Outlet
Installation
1
Loosen the union nut [B] but do not remove it.
2
Push the FEP tube [A] through the union nut [B] as far as it
reaches the stop of the threaded tube [E].
3
Tighten the union nut 1¾ rotation using an open ended spanner.
Hold Body from turning with a second wrench.
4
Put the FEP Tube into a pressure free drain with sufficient ca-
pacity.
A
B
C
FEP tube 8x6
Union nut
Compression ferrule
D
E
F
Compression cone
Threaded tube
Tightened connection
A
B C D
E
F