AMI INSPECTOR Oxygen
Program List and Explanations
A-96.250.701 / 140716
53
3.5 Quality Assurance
Performs a Quality Assurance according to your settings. Follow
the commands on the screen. Detailed explanation see
Quality
Assurance of the Instrument, p. 34
.
4 Operation
4.1 Sensors
4.1.1
Filter Time Constant:
Used to damp noisy signals. The higher the
filter time constant, the slower the system reacts to changes of the
measured value.
Range: 5–300 Sec
4.1.2
Hold after Cal:
Delay permitting the instrument to stabilize again
after calibration. During calibration plus hold-time, the signal
outputs are frozen (held on last valid value), alarm values, limits are
not active.
Range: 0–6‘000 Sec
4.2 Relay Contacts
See
Relay Contacts, p. 21
4.3 Logger
The instrument is equipped with an internal logger. The data can be
copied to the USB stick installed in the transmitter.
The logger can save approx. 1500 data records. The Records
consists of: Date, time, alarms, measuring values, raw values, case
temperature, flow.
4.3.1
Log Interval:
Select a convenient log interval. Consult the table
below to estimate the max logging time. When the logging buffer is
full, the oldest data record is erased to make room for the newest
one (circular buffer).
Range: 1 Second to 1 hour
4.3.2
Clear Logger:
If confirmed with
yes
, the complete logger data is
deleted. A new data series is started.
4.3.3
Eject USB Stick
With this function all logger data are copied to the USB stick before
the USB stick is deactivated.
Interval
1 s
5 s
1 min
5 min
10 min
30 min
1 h
Time
25 min
2 h
25 h
5 d
10 d
31 d
62 d