224
SVAN 977C User Manual - Appendixes
Preamplifier
SV 12L
nominal preamplifier attenuation: 0.7 dB;
IEPE Type
– power supply 1.5 mA@30 V
SV12L typical frequency response
Maximum peak voltage
30 V Peak-Peak (Maximum peak voltage of input sinusoidal
signal, which can be lead to the SLM without destruction the
meter)
Warm-up time
1 min. (for 0.1 dB accuracy)
Nominal delay
less than 3 seconds
(between operating of the "Reset- Button"
and beginning of a new measurement)
Typical stabilization time
after change in
environmental conditions by 20
C
1 hour
Time shift after completion
of a
measurement, before a measurement is
shown
< 1 sec
Note: When SVAN 977C is moved from a warm environment with high humidity to a colder
environment, care should be taken not to produce condensation inside the instrument. In this case,
much longer stabilization periods may be necessary.
Environmental, electrostatic
and radio frequency
criteria
Effect of humidity
< 0.5 dB (for 30%<RH<90% at 40
C and 1000 Hz)
Effect of magnetic field
< 15 dB (A) or < 25 dB (Z) (for 80 A/m and 50 Hz)
Effect of radio frequency fields
< +/-0.5 dB @ 74 dB and 10V/m
electromagnetic field
The greatest susceptibility (the least immunity) is achieved when the SLM is placed parallel to the radio
frequency field and the
Z
filter and time weighting
Fast
are selected and the SPL measurements are
considered.
The instrument produces greatest radio-frequency emission when an extension cable is connected. The cable
placed as a solenoid may produce unexpected emission depending on its physical dimensions. Any
configuration w/o extension cable reduces emission below 30 dBuV/m.
Effect of electrostatic discharge
meets requirements of IEC 61672-1:2013
During electrostatic discharge, the influence of the displayed results could be observed.
No changes in instrument operation state, configuration or stored data corruption were found out.
Effect of ambient pressure
< 0.01 dB/kPa
Summary of Contents for Shawcity SVAN 977C
Page 231: ...231 SVAN 977C User Manual Appendixes 630 Hz 800 Hz 1000 Hz 1250 Hz 1600 Hz 2000 Hz...
Page 232: ...232 SVAN 977C User Manual Appendixes 2240 Hz 2500 Hz 2800 Hz 3150 Hz 3550 Hz 4000 Hz...
Page 233: ...233 SVAN 977C User Manual Appendixes 4500 Hz 5000 Hz 5600 Hz 6300 Hz 7100 Hz 8000 Hz...
Page 234: ...234 SVAN 977C User Manual Appendixes 8500 Hz 9000 Hz 9500 Hz 10000 Hz 10600 Hz 11200 Hz...
Page 242: ...242 SVAN 977C User Manual Appendixes 250 Hz 315 Hz 400 Hz 500 Hz 630 Hz 800 Hz...
Page 243: ...243 SVAN 977C User Manual Appendixes 1000 Hz 1250 Hz 1600 Hz 2000 Hz 2240 Hz 2500 Hz...
Page 244: ...244 SVAN 977C User Manual Appendixes 2800 Hz 3150 Hz 3550 Hz 4000 Hz 4500 Hz 5000 Hz...
Page 245: ...245 SVAN 977C User Manual Appendixes 5600 Hz 6300 Hz 7100 Hz 8000 Hz 8500 Hz 9000 Hz...
Page 246: ...246 SVAN 977C User Manual Appendixes 9500 Hz 10000 Hz 10600 Hz 11200 Hz 11800 Hz 12500 Hz...
Page 250: ...250 SVAN 977C User Manual Appendixes...
Page 336: ...336 SVAN 977C User Manual Appendixes O ring gasket locations inside...