82
SVAN 977C User Manual
7.9
R
EMOTE COMMUNICATION SETTINGS
–
W
IRELESS
T
RANSFER
SVAN 977C instrument is not equipped with the modem and itself cannot assure data transfer via mobile
network. However, it can control data transfer via the modem that supports GSM connection with the help of the
Wireless Transfer
functionality. The connection with the external modem is carried out via the USB port.
Such solution is used in the SV 277 monitoring stations.
The
Wireless Transfer
position allows you
to select the network type and set the
parameters of data transmission and the
Network Reset
position to carry out the
factory setting of the network.
<ENT>
7.9.1 Selecting network type
– Network
In the
Network
screen you can disable (
Off
)
data transfer or enable it selecting
GPRS
.
<ENT>
If the
GPRS
network is selected the
Wireless Transfer
screen will have
additional positions enabling configuring GPRS connection:
Modem
,
Modem Connection
,
SMS Options
and
E-mail Settings
.
7.9.2 Configuring modem basic settings
– Modem
The
Modem
position allows you to configure
modem and data transmission settings.
<ENT>
Note: For connection with the SvanNET server, the positions in the
Modem list should be set on default values as is shown on the
screen-shots above and right-hand.
Summary of Contents for Shawcity SVAN 977C
Page 231: ...231 SVAN 977C User Manual Appendixes 630 Hz 800 Hz 1000 Hz 1250 Hz 1600 Hz 2000 Hz...
Page 232: ...232 SVAN 977C User Manual Appendixes 2240 Hz 2500 Hz 2800 Hz 3150 Hz 3550 Hz 4000 Hz...
Page 233: ...233 SVAN 977C User Manual Appendixes 4500 Hz 5000 Hz 5600 Hz 6300 Hz 7100 Hz 8000 Hz...
Page 234: ...234 SVAN 977C User Manual Appendixes 8500 Hz 9000 Hz 9500 Hz 10000 Hz 10600 Hz 11200 Hz...
Page 242: ...242 SVAN 977C User Manual Appendixes 250 Hz 315 Hz 400 Hz 500 Hz 630 Hz 800 Hz...
Page 243: ...243 SVAN 977C User Manual Appendixes 1000 Hz 1250 Hz 1600 Hz 2000 Hz 2240 Hz 2500 Hz...
Page 244: ...244 SVAN 977C User Manual Appendixes 2800 Hz 3150 Hz 3550 Hz 4000 Hz 4500 Hz 5000 Hz...
Page 245: ...245 SVAN 977C User Manual Appendixes 5600 Hz 6300 Hz 7100 Hz 8000 Hz 8500 Hz 9000 Hz...
Page 246: ...246 SVAN 977C User Manual Appendixes 9500 Hz 10000 Hz 10600 Hz 11200 Hz 11800 Hz 12500 Hz...
Page 250: ...250 SVAN 977C User Manual Appendixes...
Page 336: ...336 SVAN 977C User Manual Appendixes O ring gasket locations inside...