12
SVAN 977C User Manual
1 INTRODUCTION
SVAN 977C
is an all-digital, Class 1 Sound & Vibration level meter designed for
general acoustic and vibration, environmental noise, occupational health and
building acoustic measurements.
Three acoustic or vibration user configurable profiles allow parallel
measurements with independently defined frequency filters and RMS detector
time constants. Each profile provides significant number of results (like
Spl
,
Leq,
Sel
,
Lden
,
LEPd
,
Ltm3
,
LTeq
,
Peak
,
Max
and
Min
as well as
ten
statistics, two
rolling Leq, expected Leq value, standard Leq deviation and noise criterium and
noise ratio indexes
in case of sound measurements or
RMS
,
Ovl
,
Peak
,
P-P
in
case of vibration measurements). Advanced time history logging for each profile
provides complete information about the measured signal using the SD-card
fitted in the bottom of the meter and can be easy downloaded to any PC using
the USB interface and + software.
All required weighting filters:
A
,
B
,
C
,
Z
,
U
,
AU
and
LF
for sound measurements
and
HP1
,
HP3
,
HP10
,
Vel1
,
Vel3
,
Vel10
,
VelMF
,
Dil1
,
Dil3
,
Dil10
and
Wh
for
general vibration measurements (like acceleration, velocity and displacement);
are available with this instrument.
Using the computational power of its digital signal processor SVAN 977C can,
simultaneously to the meter mode, perform the real time
1/1 Octave
or
optionally
1/3 Octave
analysis including calculations of statistical levels
or can
optionally perform
FFT
analysis.
The built-in frequency analyser offers 1/1 octave real-time analysis as a
standard feature and 1/3 octave and FFT as optional functions. The frequency
analysis is logged together with the time history of sound or vibration meter
results.
Time domain waveform signal recording on the SD-card is available as an option
and advanced trigger and alarm functions are available in the standard version
of this instrument.
For building acoustic applications, SVAN 977C offers Reverberation Time
measurements (RT60) option working with the
BA Assistant
application for
smartphones.
SVAN 977C enables also STIPA calculations in accordance with IEC 60268-16:2011 as optional function.
A fast USB 1.1 interface (12 MHz) creates a real-time link for the PC "front-end" application of SVAN 977C. With
the use of optional interfaces (RS 232) the instrument can be remotely controlled from the PC with the use of
+
software.
SVAN 977C is equipped with
Bluetooth
®
1
v.2.0+EDR module and can be remotely control by the
SvanMOBILE
and
BA Assistant
smartphone applications.
Working as a part of the
SV 277
monitoring station, equipped with 3G modem, SVAN 977C can transfer
measured data via Internet to PC with the use of
+_RC
option or via
SvanNET
web service. The
instrument can be fully remotely controlled via these interfaces. The instrument has
advanced alarms
features,
enables the user notification about exceeded threshold levels by SMS or mails.
The instrument is powered from four AA standard alkaline or rechargeable batteries (i.e. NiMH
– a separate
charger is required). Powering the instrument from the External DC power source or the USB interface is also
possible.
1
“The
Bluetooth
®
word mark and logos are registered trademarks owned by Bluetooth SIG, Inc. and any use of such marks by
SVANTEK is under license. Other trademarks and trade names are those of their respective owners.
Summary of Contents for Shawcity SVAN 977C
Page 231: ...231 SVAN 977C User Manual Appendixes 630 Hz 800 Hz 1000 Hz 1250 Hz 1600 Hz 2000 Hz...
Page 232: ...232 SVAN 977C User Manual Appendixes 2240 Hz 2500 Hz 2800 Hz 3150 Hz 3550 Hz 4000 Hz...
Page 233: ...233 SVAN 977C User Manual Appendixes 4500 Hz 5000 Hz 5600 Hz 6300 Hz 7100 Hz 8000 Hz...
Page 234: ...234 SVAN 977C User Manual Appendixes 8500 Hz 9000 Hz 9500 Hz 10000 Hz 10600 Hz 11200 Hz...
Page 242: ...242 SVAN 977C User Manual Appendixes 250 Hz 315 Hz 400 Hz 500 Hz 630 Hz 800 Hz...
Page 243: ...243 SVAN 977C User Manual Appendixes 1000 Hz 1250 Hz 1600 Hz 2000 Hz 2240 Hz 2500 Hz...
Page 244: ...244 SVAN 977C User Manual Appendixes 2800 Hz 3150 Hz 3550 Hz 4000 Hz 4500 Hz 5000 Hz...
Page 245: ...245 SVAN 977C User Manual Appendixes 5600 Hz 6300 Hz 7100 Hz 8000 Hz 8500 Hz 9000 Hz...
Page 246: ...246 SVAN 977C User Manual Appendixes 9500 Hz 10000 Hz 10600 Hz 11200 Hz 11800 Hz 12500 Hz...
Page 250: ...250 SVAN 977C User Manual Appendixes...
Page 336: ...336 SVAN 977C User Manual Appendixes O ring gasket locations inside...