214
SVAN 977C User Manual - Appendixes
B.2.1.1. Record with the results
The contents of the record with the results depends on the selected measurement function and the value set
in the
LOGGER
position of the
PROFILE x
and
SPECTRUM
sub-lists. The following elements can be present
(in the given sequence):
1. flag record
<flags> - b0: 1- the overload detected, 0 - the overload not detected
2. results of the measurement from the first profile if the corresponding
LOGGER
position was active
(BufferP [1] in 0 or 0); up to six words are written:
SLM mode:
<result1> -
Peak
result, depending on the value of BufferP[1] (cf. Table B.1.12)
<result2> -
Max
result, depending on the value of BufferP[1] (cf. Table B.1.12)
<result3> -
Min
result, depending on the value of BufferP[1] (cf. Table B.1.12)
<result4> -
Leq
result, depending on the value of BufferP[1] (cf. Table B.1.12)
<result5> -
LR1
result, depending on the value of BufferP[1] (cf. Table B.1.12)
<result6> -
LR2
result, depending on the value of BufferP[1] (cf. Table B.1.12)
VLM mode:
<result1> -
Peak
result, depending on the value of BufferP[1] (cf. Table B.1.13)
<result2> -
P-P
result, depending on the value of BufferP[1] (cf. Table B.1.13)
<result3> -
Max
result, depending on the value of BufferP[1] (cf. Table B.1.13)
<result4> -
RMS
result, depending on the value of BufferP[1] (cf. Table B.1.13)
3. results of the measurement from the second profile if the corresponding
LOGGER
position was active
(BufferP [2] in Table B.1.12 or Table B.1.13); up to six words are written:
SLM mode:
<result1> -
Peak
result, depending on the value of BufferP[2] (cf. Table B.1.12)
<result2> -
Max
result, depending on the value of BufferP[2] (cf. Table B.1.12)
<result3> -
Min
result, depending on the value of BufferP[2] (cf. Table B.1.12)
<result4> -
Leq
result, depending on the value of BufferP[2] (cf. Table B.1.12)
<result5> -
LR1
result, depending on the value of BufferP[2] (cf. Table B.1.12)
<result6> -
LR2
result, depending on the value of BufferP[2] (cf. Table B.1.12)
VLM mode:
<result1> -
Peak
result, depending on the value of BufferP[2] (cf. Table B.1.13)
<result2> -
P-P
result, depending on the value of BufferP[2] (cf. Table B.1.13)
<result3> -
Max
result, depending on the value of BufferP[2] (cf. Table B.1.13)
<result4> -
RMS
result, depending on the value of BufferP[2] (cf. Table B.1.13)
(4) results of the measurement from the third profile if the corresponding
LOGGER
position was active
(BufferP [3] in Table B.1.12 or Table B.1.13); up to six words are written:
SLM mode:
<result1> -
Peak
result, depending on the value of BufferP[3] (cf. Table B.1.12)
<result2> -
Max
result, depending on the value of BufferP[3] (cf. Table B.1.12)
<result3> -
Min
result, depending on the value of BufferP[3] (cf. Table B.1.12)
<result4> -
Leq
result, depending on the value of BufferP[3] (cf. Table B.1.12)
<result5> -
LR1
result, depending on the value of BufferP[3] (cf. Table B.1.12)
<result6> -
LR2
result, depending on the value of BufferP[3] (cf. Table B.1.12)
Summary of Contents for Shawcity SVAN 977C
Page 231: ...231 SVAN 977C User Manual Appendixes 630 Hz 800 Hz 1000 Hz 1250 Hz 1600 Hz 2000 Hz...
Page 232: ...232 SVAN 977C User Manual Appendixes 2240 Hz 2500 Hz 2800 Hz 3150 Hz 3550 Hz 4000 Hz...
Page 233: ...233 SVAN 977C User Manual Appendixes 4500 Hz 5000 Hz 5600 Hz 6300 Hz 7100 Hz 8000 Hz...
Page 234: ...234 SVAN 977C User Manual Appendixes 8500 Hz 9000 Hz 9500 Hz 10000 Hz 10600 Hz 11200 Hz...
Page 242: ...242 SVAN 977C User Manual Appendixes 250 Hz 315 Hz 400 Hz 500 Hz 630 Hz 800 Hz...
Page 243: ...243 SVAN 977C User Manual Appendixes 1000 Hz 1250 Hz 1600 Hz 2000 Hz 2240 Hz 2500 Hz...
Page 244: ...244 SVAN 977C User Manual Appendixes 2800 Hz 3150 Hz 3550 Hz 4000 Hz 4500 Hz 5000 Hz...
Page 245: ...245 SVAN 977C User Manual Appendixes 5600 Hz 6300 Hz 7100 Hz 8000 Hz 8500 Hz 9000 Hz...
Page 246: ...246 SVAN 977C User Manual Appendixes 9500 Hz 10000 Hz 10600 Hz 11200 Hz 11800 Hz 12500 Hz...
Page 250: ...250 SVAN 977C User Manual Appendixes...
Page 336: ...336 SVAN 977C User Manual Appendixes O ring gasket locations inside...