259
SVAN 977C User Manual - Appendixes
Basic accuracy
<
0.2
dB (for the temperature T=+23°C
5°C for sinusoidal
signal 1 V
RMS
in the band 10
Hz ÷ 20 kHz with
HP
filter)
Measurement error
in the full temperature range
<
0.1 dB (when the temperature is from -
10°C to +50°C for the
sinusoidal signal 1 V
RMS
in the band 10
Hz ÷ 20 kHz with
Z
filter)
Overload detector
The instrument has the built-in overload detectors. Both A/D converter and input amplifier overload conditions
are detected. The overload in the measurement channel (in its analogue part) and the overload of the
analogue /
digital converter are both detected. The “overload” indication appears when the input signal
amplitude is 0.5 dB above t
he declared “Peak measurement range”.
Warm-up time
1 min. (for 0.1 dB accuracy).
Effect of humidity
< 0.5 dB (for 30%<RH<90% at 40
C re Reference conditions)
Effect of magnetic field
< 5 µV
RMS
(A) or < 20
µV
RMS
(Z) (for 80 A/m and 50 Hz)
Effect of Vibration
< 0.1 dB (from 20 Hz to 1000 Hz at 1 m/s
2
)
Electrostatic
and radio frequency
criteria
– see Chapter C.1.1
Antialiasing filter
Built-in antialiasing filter. Second-order analogue filter, passive Class, combined with on-chip FIR digital filter
of the analogue-to-digital converter, ensuring correct sampling of the measured signal.
Pass band (-1 dB)
22.200 kHz or 44.4 kHz
Pass band (-3 dB)
23.520 kHz or 47.0 kHz
Stop band
26.256 kHz or 52.53 kHz
Attenuation in the stop band
> 80 dB.
Sampling frequency
48 kHz or 96 kHz
(internal only).
Analogue to digital converter
1 x 24 bit
Reference range
Low
Input attenuator accuracy
0.1 dB (for f = 1
kHz and T = +23°C)
Internal oscillator accuracy
0.01 % (for f = 1
kHz and T = +23°C)
Digital Filters
Weighting filters
•
A meeting requirements of the IEC 61672-1:2013 standard for the Class
1 “A” filter
•
C meeting requirements of the IEC 61672-1:2013 standard for the Class
1 “C” filter
•
Z meeting requirements of the IEC 61672-1:2013 standard for the Class
1 “Z“ filter
•
B meeting IEC651 for the Class 1 filter
See Chapter C.4 for the A, C, B and Z filters characteristics.
Noise voltage measured with SV12L preamplifier, equivalent impedance-adapter Class of ST02 and 50
Ω
input impedance, 20 kHz Bandwidth.
Summary of Contents for Shawcity SVAN 977C
Page 231: ...231 SVAN 977C User Manual Appendixes 630 Hz 800 Hz 1000 Hz 1250 Hz 1600 Hz 2000 Hz...
Page 232: ...232 SVAN 977C User Manual Appendixes 2240 Hz 2500 Hz 2800 Hz 3150 Hz 3550 Hz 4000 Hz...
Page 233: ...233 SVAN 977C User Manual Appendixes 4500 Hz 5000 Hz 5600 Hz 6300 Hz 7100 Hz 8000 Hz...
Page 234: ...234 SVAN 977C User Manual Appendixes 8500 Hz 9000 Hz 9500 Hz 10000 Hz 10600 Hz 11200 Hz...
Page 242: ...242 SVAN 977C User Manual Appendixes 250 Hz 315 Hz 400 Hz 500 Hz 630 Hz 800 Hz...
Page 243: ...243 SVAN 977C User Manual Appendixes 1000 Hz 1250 Hz 1600 Hz 2000 Hz 2240 Hz 2500 Hz...
Page 244: ...244 SVAN 977C User Manual Appendixes 2800 Hz 3150 Hz 3550 Hz 4000 Hz 4500 Hz 5000 Hz...
Page 245: ...245 SVAN 977C User Manual Appendixes 5600 Hz 6300 Hz 7100 Hz 8000 Hz 8500 Hz 9000 Hz...
Page 246: ...246 SVAN 977C User Manual Appendixes 9500 Hz 10000 Hz 10600 Hz 11200 Hz 11800 Hz 12500 Hz...
Page 250: ...250 SVAN 977C User Manual Appendixes...
Page 336: ...336 SVAN 977C User Manual Appendixes O ring gasket locations inside...