50
SVAN 977C User Manual
In the time edition line, you may define hour
and minute of the split operation and switch
on (
) or off (
) the current split.
After pressing
<ENTER>
the
Split Time X
list will show the active times of the day
when the logger will be split.
<ENT>
4.4.2 Selecting results for logging
– Logger Results
In the
Logger Results
list, you can select
results for three independent profiles (
Peak
,
Max
,
Min
,
Leq
,
LXY
,
LR1
and
LR2
for
Sound measurements and
Peak
,
P-P
,
Max
and
RMS
for Vibration measurements) and
results taken from the weather station
(
Meteo
), which will be logged in a logger file
with the
Logger Step
.
Activation / deactivation can be done with
the
◄ / ► key pressed together with
<Alt>
.
<ENT>
Note: When Logger is disabled or no Logger results were selected, the logger plot cannot be
activated in Display Modes
and therefore doesn’t appear on the display.
When
Logger
is enabled and results for
logging selected, the logger plot will be
available in measurement views.
<Alt/
▲
>
4.4.3 Selecting additional summary results for saving
– Summary Results
The
Summary Results
list allows you to
include/exclude logging of some summary
results,
like:
Statistics
,
Meteo
,
Statistics Spectrum
,
Max Spectrum
and
Min Spectrum
for Sound measurements
and
Meteo
,
Max Spectrum
and
Min Spectrum
for Vibration measurements.
Default modes: all enabled.
<ENT>
Note: Meteo results are transmitted to the instrument from the weather station connected to the
SV 277 monitoring station.
4.4.4 Configuring Logger trigger
– Logger Trigger
In the
Logger Trigger
list, you can
configure the way the logger results will be
registered in the logger file. It is a context list
in which: the trigger can be disabled (
Off
) or
enabled by selecting its type (
Level+
or
Level-
) in the
Trigger
position. Default
value:
Off
.
<ENT>
Summary of Contents for Shawcity SVAN 977C
Page 231: ...231 SVAN 977C User Manual Appendixes 630 Hz 800 Hz 1000 Hz 1250 Hz 1600 Hz 2000 Hz...
Page 232: ...232 SVAN 977C User Manual Appendixes 2240 Hz 2500 Hz 2800 Hz 3150 Hz 3550 Hz 4000 Hz...
Page 233: ...233 SVAN 977C User Manual Appendixes 4500 Hz 5000 Hz 5600 Hz 6300 Hz 7100 Hz 8000 Hz...
Page 234: ...234 SVAN 977C User Manual Appendixes 8500 Hz 9000 Hz 9500 Hz 10000 Hz 10600 Hz 11200 Hz...
Page 242: ...242 SVAN 977C User Manual Appendixes 250 Hz 315 Hz 400 Hz 500 Hz 630 Hz 800 Hz...
Page 243: ...243 SVAN 977C User Manual Appendixes 1000 Hz 1250 Hz 1600 Hz 2000 Hz 2240 Hz 2500 Hz...
Page 244: ...244 SVAN 977C User Manual Appendixes 2800 Hz 3150 Hz 3550 Hz 4000 Hz 4500 Hz 5000 Hz...
Page 245: ...245 SVAN 977C User Manual Appendixes 5600 Hz 6300 Hz 7100 Hz 8000 Hz 8500 Hz 9000 Hz...
Page 246: ...246 SVAN 977C User Manual Appendixes 9500 Hz 10000 Hz 10600 Hz 11200 Hz 11800 Hz 12500 Hz...
Page 250: ...250 SVAN 977C User Manual Appendixes...
Page 336: ...336 SVAN 977C User Manual Appendixes O ring gasket locations inside...