SVAN 977C User Manual
Gradient trigger
The
Grad +
trigger starts the 1-second measurement/integration on condition:
value of the measured result (
Source
) averaged by 0.5 ms is greater than the
threshold value (
Level
) and the gradient of this Source is greater than the
gradient threshold value (
Gradient
). In other cases, the instrument continues
checking the trigger condition every 0.5 ms.
This trigger has the same logic as the
Level +
trigger, but the trigger condition requires also the gradient threshold
level to be exceeded.
External trigger
When
External
is selected, the measurement is triggering by the external signal
on the
I/O
socket. In this case, it is necessary to set up the
I/O Mode
parameter
as
Digital In
(
path:
<
Menu> / Instrument / Multifunction I/O
).
After the measurement/integration start from the external trigger, the
measurement/integration will continue for the
Integration Period
.
Note: When a measurement is waiting for the gradient or external
trigger the “trigger” icon appears alternatively with the „measurement”
icon.
Source result
Only one measured result (
Source
) can be used for checking trigger condition in the
Level Meter
mode, namely
the instantaneous RMS from the first profile (with appropriate filter and detector), which is denoted here as
Leq(1)
in the sound mode and
RMS(1)
in the vibration mode. This position cannot be changed.
Threshold level
The threshold level for the measurement trigger
(
Level
) can be set with 1 dB step in the range
from 24 dB to 136 dB in sound modes and from
64dB (1.585mm/s
2
) to 176dB (631 m/s
2
) in
vibration modes.
Rate of Source value changing
This position appears when the
Grad+
trigger is chosen. The rate of the Source
value changing (
Gradient
) can be set in the range from 1 dB/ms to 100 dB/ms.
4.3 S
ETTING PARAMETERS FOR PROFILES
–
P
ROFILES
The
Profiles
screen enables programming
of three virtual level meters (
Profiles
), which
measure the set of results with the same
weighting filter (
Filter
) and exponential
detector time constant (
Detector
) (see
Appendix D). Default values
– see
<ENT>
Summary of Contents for Shawcity SVAN 977C
Page 231: ...231 SVAN 977C User Manual Appendixes 630 Hz 800 Hz 1000 Hz 1250 Hz 1600 Hz 2000 Hz...
Page 232: ...232 SVAN 977C User Manual Appendixes 2240 Hz 2500 Hz 2800 Hz 3150 Hz 3550 Hz 4000 Hz...
Page 233: ...233 SVAN 977C User Manual Appendixes 4500 Hz 5000 Hz 5600 Hz 6300 Hz 7100 Hz 8000 Hz...
Page 234: ...234 SVAN 977C User Manual Appendixes 8500 Hz 9000 Hz 9500 Hz 10000 Hz 10600 Hz 11200 Hz...
Page 242: ...242 SVAN 977C User Manual Appendixes 250 Hz 315 Hz 400 Hz 500 Hz 630 Hz 800 Hz...
Page 243: ...243 SVAN 977C User Manual Appendixes 1000 Hz 1250 Hz 1600 Hz 2000 Hz 2240 Hz 2500 Hz...
Page 244: ...244 SVAN 977C User Manual Appendixes 2800 Hz 3150 Hz 3550 Hz 4000 Hz 4500 Hz 5000 Hz...
Page 245: ...245 SVAN 977C User Manual Appendixes 5600 Hz 6300 Hz 7100 Hz 8000 Hz 8500 Hz 9000 Hz...
Page 246: ...246 SVAN 977C User Manual Appendixes 9500 Hz 10000 Hz 10600 Hz 11200 Hz 11800 Hz 12500 Hz...
Page 250: ...250 SVAN 977C User Manual Appendixes...
Page 336: ...336 SVAN 977C User Manual Appendixes O ring gasket locations inside...