6.2 Precautions
CAUTION: Personal Injury or Equipment Damage Hazard.
Safety precautions must
always be exercised when using electrical equipment to prevent user/patient shock, fire
hazard and equipment damage
.
CAUTION: Personal Injury or Equipment Damage Hazard.
Federal (USA) law restricts
this device to sale by or on the order of a physician.
CAUTION: Interference Hazard.
Portable and mobile RF communications equipment can
affect medical electrical equipment. Electromagnetic waves emitted by these types of
equipment may result in equipment failure or the gi4000 ESU not performing properly.
→
Reference section
8.12
Technical
Specifications:
Standards on
Electromagnetic Compatibility (EMC)
of this manual for manufacturer
recommended separation distances between portable and mobile RF
communications equipment and the gi4000 ESU.
CAUTION: Equipment Damage Hazard.
Unsuitable operating temperature and humidity
may cause equipment damage or failure. Reference section
8.4 Technical Specifications:
Environmental Operating Parameters
for manufacturer recommended operating
parameters.
CAUTION: Equipment Damage Hazard.
Unsuitable transport or storage temperature and
humidity may cause equipment damage or failure. Reference section
8.5 Technical
Specifications: Environmental Transport and Storage Parameters
for manufacturer
recommended Transport and Storage parameters.
→
If the equipment was transported or stored below or above the recommended
temperature, pressure and humidity levels, the equipment will require a certain
acclimation time at a specified temperature range. Insufficient acclimatization
time or temperature may cause equipment damage or failure. Reference
section
8.5 Technical Specifications: Environmental Transport and
Storage Parameters
for manufacturer recommended acclimatization
parameters.
CAUTION: Equipment Damage Hazard.
Install the device in such a way that there is
unobstructed circulation of air around the housing. If ventilation is unsuitable, the device
can overheat and may cause equipment damage or failure.
CAUTION:
Equipment Damage Hazard.
Precaution must be taken when handling liquids
around electrical equipment. The gi4000 ESU is not water tight; therefore, liquid penetration
may result in equipment failure or the gi4000 ESU not performing properly.
→
DO NOT
operate the gi4000 ESU if liquid has been spilled on the unit.
CAUTION: Equipment Damage Hazard. NEVER
place the gi4000 ESU on top of other
electrical equipment due to its fluid handling system.
→
DO NOT
place the sterile water bottle on top of the gi4000 ESU. Utilize the
Water Bottle Holder provided for securing water bottles during use.
Summary of Contents for G1110001
Page 1: ...gi4000 Electrosurgical Unit Reference Reorder No G1110001 User and Maintenance Manual ...
Page 6: ... This page is intentionally left blank ...
Page 54: ...00732467 Rev D Page 48 of 110 8 11 9 Bipolar Method Figure 8 25 Figure 8 26 Figure 8 27 ...
Page 109: ...00732467 Rev D Page 103 of 110 Figure 13 21 Argon RF Output Test Diagram ...