00732467 Rev. D
Page
35
of
110
8.0 Technical Specifications
8.1 General
Internal Design:
Solid State
Output Configuration:
Isolated (RF Floating)
Cooling:
Natural Convection, Rear Panel Vent
Mounting:
GoCart Mobile Cart or a flat, stable surface which
can withstand the weight of the gi4000 ESU.
Reference section
8.3 Dimensions and Weight
.
Service Life:
7 Years
Equipotential Ground Connection:
An equipotential ground connection is provided to
allow connection of the gi4000 ESU to a common
ground by way of a potential equalization connector.
This connection meets IEC 60601-1 requirements.
8.2 Classifications
Type of Protection Against
Electrical Shock:
Class I (IEC 60601-1)
Degree of Protection Against
Electrical Shock:
Type CF Equipment (IEC 60601-1)
Defibrillation Proof Applied Parts
Applied parts are the electrosurgical accessories
connected to the output receptacles of the unit,
including but not limited to: dispersive electrodes,
ArConnect® argon probe connector,
ArC Smart™
argon coagulation probe(s), APC probes, monopolar
active
cords,
hot
snares,
hot
forceps,
sphincterotomes, TouchSoft Coagulator® Monopolar
Probe, bipolar probes, irrigation tubing, etc.
Degree of Ingress Protection
(IP Rating):
IPX0 (gi4000 ESU)
IP68 (gi4000 ESU footswitch)
Sterilization:
The gi4000 ESU and accessories are
NOT
intended
to be sterilized or autoclaved.
Summary of Contents for G1110001
Page 1: ...gi4000 Electrosurgical Unit Reference Reorder No G1110001 User and Maintenance Manual ...
Page 6: ... This page is intentionally left blank ...
Page 54: ...00732467 Rev D Page 48 of 110 8 11 9 Bipolar Method Figure 8 25 Figure 8 26 Figure 8 27 ...
Page 109: ...00732467 Rev D Page 103 of 110 Figure 13 21 Argon RF Output Test Diagram ...