13.5 Fuse Replacement
1. Switch the power ON/OFF s
witch to the OFF position (denoted by “O” on the switch,
switch should not be illuminated).
2. Disconnect the power cord from the wall outlet and the rear panel of the unit.
3. Depress the tabs located on the left and right sides of the fuse holder located on the
rear panel of the unit to remove the fuse holder from the power cord receptacle (see
Figure 13-32
).
4. Replace the fuses as marked on the rear panel of the unit (see
Figure 13-33
).
5. Holding the unit in place, reinstall the fuse holder into the power cord receptacle. The
fuse h
older will “click” into place.
6. Reconnect the power cord to the rear panel of the unit and the wall outlet.
7. Switch the power ON/OFF s
witch to the ON position (denoted by the “I”, switch will
illuminate).
8. Visually verify the power ON/OFF switch illuminates.
Figure 13-30
Figure 13-31
Figure 13-32
Figure 13-33