00732467 Rev. D
Page
52
of
110
The gi4000 ESU is intended for use in an electromagnetic environment in which radiated RF
disturbances are controlled. The user can help prevent electromagnetic interference by
maintaining a minimum distance between portable and mobile RF communications equipment
(transmitters) and the gi4000 ESU as recommended below, according to the maximum output
power of the communications equipment.
Table 8-4
R
ATED MAXIMUM OUTPUT POWER OF
TRANSMITTER
S
EPARATION DISTANCE ACCORDING TO FREQUENCY OF
TRANSMITTER
150
K
H
Z TO
80
MH
Z
P
d
=
3
5
.
3
80
MH
Z TO
800
MH
Z
P
d
=
3
5
.
3
800
MH
Z TO
2.5
GH
Z
P
d
=
3
7
0.01 W
0.12 m
0.12 m
0.23 m
0.1 W
0.37 m
0.37 m
0.74 m
1 W
1.17 m
1.17 m
2.33 m
10 W
3.69 m
3.69 m
7.38 m
100 W
11.67 m
11.67 m
23.33 m
For transmitters rated at a maximum output power not listed above, the recommended separation distance d in meters (m)
can be determined using the equation applicable to the frequency of the transmitter, where P is the maximum output power
rating of the transmitter in watts (W) according to the transmitter manufacturer.
NOTE 1 At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection
from structures, objects and people.
Summary of Contents for G1110001